ISO 11040-4
Prefilled Glass Syringe Testing: Break-Loose, Glide Force and Breakage

ISO 11040-4 prefilled glass syringe test fixture with guided ram by Universal Grip

Downloads: ISO 11040-4 Prefilled Glass Syringe Test Fixture – Universal Grip Data Sheet (PDF)

Written by Chris Thomas, Universal Grip Company · Reviewed by Frank Bertini · Updated October 31, 2026

ISO 11040-4 is the international standard for glass barrels for injectables and sterilized subassembled syringes ready for filling, the nested, pre-sterilized glass syringes that pharmaceutical companies fill with drug product. It sets materials, dimensions, quality and performance requirements, and its annexes give the mechanical test methods used to prove them: finger flange and front-end (Luer cone) breakage resistance, glide force, needle penetration, needle pull-out, Luer lock collar and tip cap tests, and closure tests. Syringe manufacturers, fill-finish sites, combination-product developers and contract test labs use these methods for design verification, incoming inspection and release testing. They are easy to get wrong. The 2024 edition renamed and tightened several requirements, the glide force method is often confused with break-loose and extrusion testing of the finished drug product in ISO 11040-8, and a few degrees of off-axis loading on a glass barrel can change both force curves and breakage results. This guide explains what the current edition covers, how the axial plunger tests work, how to evaluate the force curve with a worked example, and how to set the tests up with the Universal Grip prefilled glass syringe fixture.

ISO 11040-4 at a glance

  • Current edition: ISO 11040-4:2024, the fourth edition, published June 2024 by ISO/TC 76. It cancels and replaces ISO 11040-4:2015.

  • Scope: materials, dimensions, quality and performance requirements, and test methods, for tubing-glass barrels (single-chamber) for injection preparations and for sterilized subassembled syringes ready for filling. These are single-use products. Plunger stoppers and plunger rods are outside its scope.

  • Barrel sizes: the dimension table covers nominal volumes from 0.5 mL to 20 mL.

  • Front ends: 6 % Luer slip, Luer lock adapter and staked needle designs. Luer connections must meet ISO 80369-7 and ISO 80369-1. Staked needles reference ISO 7864 and ISO 9626.

  • Test annexes (2024 edition): Annex C (normative) test methods for syringe barrels; Annex E (informative) glide force; Annex F (informative) needle penetration; Annex G (normative) test methods for front-end components; Annex H (informative) tightness test.

  • Test conditions: ambient laboratory conditions unless agreed otherwise. Test equipment must be qualified and test methods validated.

  • Number of samples: most annexes do not fix a sample size. Set it with a statistical sampling plan agreed between supplier and customer.

  • Results: peak forces (breakage, pull-off, pull-out, penetration), glide force profiles (maximum and mean), torques and pass/fail leakage checks.

  • Related standards: ISO 11040-5, -6, -7 and -8, ISO 7886-1, ISO 80369-7, ISO 80369-20, ISO 7864, ISO 9626.

What changed in ISO 11040-4:2024

The 2024 revision kept the overall structure but changed several things test labs need to know.

  • Minimum breaking forces. Finger flange breakage resistance and front-end breakage resistance now have specified minimum force values. In 2015 they were test methods without fixed limits.

  • Front-end breakage replaces Luer cone breakage. The former Luer cone breakage requirement is now called front-end breakage resistance.

  • Needle penetration limit. A maximum force for a needle to penetrate the test foil has been added, to catch needle-point and lubrication defects.

  • Glide force for non-lubricated syringes. The glide force method previously applied only to lubricated syringes. It now covers non-lubricated syringes too.

  • Staked-needle syringes. Needle diameter, needle length and finger flange design are agreed between manufacturer and customer, with a new tolerance table for volume and needle-length combinations.

  • Quality system. Testing and documentation must sit inside a formal quality management system, with qualified equipment and validated methods.

  • Annexes. Former Annex B content moved into Clause 5.1. The new Annex B shows typical components of a subassembled syringe, and the figures in Annex A were updated.

The numeric minimum and maximum values are in the standard itself. Buy the current edition and check them before writing acceptance criteria. Values quoted in older test reports were written against the 2015 edition, which had none.

What the ISO 11040-4 functional tests measure

A prefilled glass syringe has to survive filling lines, autoinjector assembly, shipping and use without breaking, leaking or jamming, and it has to deliver its dose with a predictable force. The ISO 11040-4 annexes test each mechanical interface:

Glide force and break-loose force (Annex E)

The plunger stopper is pushed through the barrel at a constant speed and the force is recorded against travel. The curve shows two regions. The break-loose force is the initial peak needed to start the stopper moving, where static friction and any stopper-to-glass adhesion are overcome. The glide force is the force needed to keep it moving, evaluated over the glide region as a maximum and a mean. A high break-loose peak or a rising, jagged glide region usually means poor or uneven siliconization, silicone migration during storage or a stopper and barrel mismatch. In ISO 11040-4 the test checks the barrel and its lubrication with water or another agreed liquid. Break-loose and extrusion forces of the finished drug-filled syringe are covered by ISO 11040-8, whose 2026 revision added a dedicated annex for that test.

Finger flange breakage resistance (Annex C)

The barrel is supported under its finger flange in a holder and an axial load is applied until the flange breaks. The peak force is the flange breakage force. The flange is the part that carries the user's fingers and the autoinjector's reaction loads, so this test matters for autoinjector programs. The 2024 edition sets a minimum value.

Front-end (Luer cone) breakage resistance (Annex C)

A lateral load is applied near the tip of the cone or staked-needle front end until it breaks, simulating side loads during needle or connector attachment. The peak force is reported and compared with the 2024 minimum.

Needle tests (Annexes F and G)

The needle penetration test measures the force for a staked needle to pierce a standard test foil. It reflects point geometry and needle lubrication, and the 2024 edition caps it with a maximum force. The needle pull-out test grips the needle and pulls it axially from the barrel to check the bond. The minimum pull-out forces by needle gauge come from ISO 7864.

Luer lock collar and closure tests (Annex G)

For Luer lock syringes, the collar is tested for pull-off force (axial) and torque resistance (rotation). The rigid tip cap unscrewing torque and the pull-off force of the tip cap or needle shield are measured as well. A closure liquid-leakage check confirms the tip cap holds while the syringe is pressurized through the plunger.

Samples: preparation and number

Test barrels and subassembled syringes as supplied: sterilized and nested, with the agreed tip cap or needle shield fitted. For glide and break-loose testing the syringe is filled with water or another agreed liquid and stoppered, by vent-tube or vacuum stoppering, with the stopper type agreed between manufacturer and customer. The plunger rod is then fitted. Record:

  • Barrel size and format (Luer slip, Luer lock or staked needle), lot and nesting position if relevant.

  • Lubrication type: sprayed silicone oil, baked-on silicone or non-lubricated.

  • Stopper type, fill liquid and fill volume.

  • Stoppering method and the time between stoppering and test. Break-loose force often rises with storage time as silicone redistributes, so time zero and aged results are not interchangeable.

  • Laboratory temperature. ISO 11040-4 calls for ambient lab conditions unless agreed otherwise.

ISO 11040-4 does not fix a sample size for most annexes. Published studies of 2015-edition methods have used 50 syringes per test, and the authors recommended at least 30 for reliable statistics on brittle glass breakage. For lot release, use an agreed sampling plan, such as an attributes plan to ISO 2859-1 or a variables plan to ISO 3951-1.

Test speed and machine requirements

Each annex defines how the load is applied and its test speed. Published work against the 2015 annexes used 100 mm/min for finger flange breakage and glide force, 25 mm/min for lateral Luer cone breakage, 50 mm/min for needle pull-out and 280 mm/min for closure-cap pull-off. Glide force is often agreed at other speeds, such as 50 mm/min, to match a device's delivery rate. Confirm each speed against the 2024 edition before validating a method, because the annexes were revised.

The forces are small:

  • Glide and break-loose forces for lubricated 1 mL syringes are typically a few newtons. A 50 N measuring range is commonly recommended for glide testing.

  • Needle pull-out minimums from ISO 7864 are in the tens of newtons.

  • Flange and front-end breakage are the highest forces in the set. Size the load cell from a few trial breaks and the 2024 minimum values for your barrel.

What matters more than capacity is resolution and speed of capture. A break-loose peak lasts a fraction of a second, so sample at 500 Hz or faster. Axial alignment matters too: a ram that is even slightly off the syringe axis adds friction and side load, inflating glide force and causing early flange or barrel cracks.

The Universal Grip ISO 11040-4 prefilled glass syringe fixture

Our Prefilled Glass Syringe Test Fixture, part ref. UGC-ISO11040-4-PF, is a vertical fixture for glass prefilled syringes. The barrel sits in an interchangeable PTFE insert and is held by a lever clamp, while a guided ram on the crosshead applies axial force for functional tests such as break-loose and gliding force. A clear chamber collects expelled liquid.

  • Guided ram keeps the force exactly on the syringe axis, which is the main source of error in glide force testing.

  • Interchangeable PTFE inserts fit a range of barrel sizes, for prefilled glass syringes from 0.5 to 10 mL.

  • Lever clamp secures the syringe quickly and repeatably, which matters when you test 30 or 50 syringes per lot.

  • Clear collection chamber contains the expelled liquid and keeps it off the load cell and frame.

  • Load capacity: 500 N standard, with other capacities on request.

  • Materials: stainless steel body and guided steel ram, with PTFE-lined barrel seats.

  • Loading: axial compression via the guided ram.

  • Machine connection: clevis pin or threaded adapter to suit your frame. Adapters are supplied, and see our adapters for other frames.

  • Compatible frames: electromechanical universal test machines and low-force frames.

The fixture is built around the axial plunger tests: break-loose and glide force, and other tests where a ram pushes along the syringe axis. ISO 11040-4 also includes tests with different load paths. Front-end breakage is a lateral load, needle and collar pull-off are tensile, and collar and tip-cap tests are torque. Those need different holders or a torque instrument. When you order, tell us the syringe sizes and formats, the annex tests you need to run, the load capacity and your test machine connection, and we will confirm what the fixture covers and what else you need. Browse related compression fixtures and puncture fixtures, or contact us. Specifications are typical and are confirmed at time of quote.

Choosing a testing machine for ISO 11040-4

Mark-10 motorized testers: a natural fit

ISO 11040-4 forces sit squarely in Mark-10 territory, and the speeds used for these annexes, from 25 mm/min to 280 mm/min, fall inside the F-series standard range of 13–330 mm/min. With ±0.2 % speed accuracy and 5,000 Hz sampling, the break-loose peak is captured with plenty of resolution.

  • The F105 (100 lbF, about 450 N) is the obvious choice for glide and break-loose force QC. Pair it with a low-capacity load cell or force gauge suited to forces of a few newtons.

  • The F305 (300 lbF, about 1.3 kN) adds headroom for flange breakage on larger barrels while still covering glide testing. The fixture's standard rating is 500 N, so order a higher capacity if you expect to exceed it.

  • Both have 457 mm of travel, more than any syringe stroke needs. See all tensile testers and force gauges.

Galdabini universal testing machines for central labs

A Galdabini 2.5 kN benchtop UTM makes sense for a central QC or development lab that runs ISO 11040-4 alongside packaging, polymer or device testing on one frame, with methods, curve-window evaluation and audit trails in software. For glide force at a few newtons, make sure the frame carries a load cell sized for that range. A 2.5 kN cell alone is too coarse. Ask us about load cell options when you specify the system. A larger frame adds nothing for these tests.

Extensometers and chambers

You do not need an extensometer. Plunger travel is measured by crosshead displacement, which is all the glide force evaluation needs. An environmental chamber is not required either, because the standard calls for ambient conditions. Some development programs agree on cold-chain or warm testing, and a chamber can support that by agreement.

Glide and break-loose force procedure, step by step

  • Prepare the syringes. Fill, stopper and fit the plunger rods as agreed, and record the stoppering time. Remove the tip cap or needle shield if the agreed method requires an open front end.

  • Set up the fixture. Fit the PTFE insert for the barrel size, mount the fixture on the base and the guided ram on the crosshead with the supplied adapter, and check the ram is centred on the plunger rod.

  • Load the syringe. Seat the barrel flange in the insert, close the lever clamp and fit the collection chamber.

  • Zero the load cell with the ram just clear of the plunger rod.

  • Approach. Bring the ram down to touch the plunger rod with a small preload, then zero displacement.

  • Run the test at the agreed speed, recording force against displacement at 500 Hz or faster, until the stopper reaches the end of its stroke. Stop on a force limit so the stopper is not driven hard into the shoulder of the barrel.

  • Evaluate the curve using the agreed windows (see below).

  • Inspect the barrel for cracks and the stopper for damage, and note any stick-slip or noise.

  • Repeat for the full sample, then summarize with mean, standard deviation and any statistical acceptance test.

Calculations with a worked example

Glide force evaluation depends on how the curve is split into windows, and the published literature notes that the glide region boundaries need prior agreement between customer and test lab. A common approach:

  • Break-loose force (F_BL): the maximum force in an initial window, for example the first 2 mm of travel.

  • Maximum glide force (F_max): the highest force in the glide window, from the end of the break-loose window to a point a few millimetres before the end of the stroke, for example 3 mm.

  • Mean glide force (F_avg): the arithmetic mean of all force readings in the glide window.

  • Pressure equivalent (optional): p = F / A, where A is the stopper cross-sectional area from the barrel inner diameter.

Worked example. A 1 mL long glass syringe filled with water is tested at 100 mm/min with a 30 mm stroke. The break-loose window is 0–2 mm and the glide window is 2–27 mm.

  • Break-loose force: the highest reading in 0–2 mm is 4.2 N at 0.6 mm of travel.

  • Glide window size: 25 mm at 100 mm/min takes 25/100 = 0.25 min = 15 s. At 5,000 Hz that is 15 × 5,000 = 75,000 readings.

  • Mean glide force: the mean of those readings is 2.1 N. Maximum glide force is 2.6 N.

  • Pressure check: with an assumed inner diameter of 6.35 mm, A = π × (6.35/2)² = π × 3.175² = 31.67 mm². The mean glide pressure is 2.1 / 31.67 = 0.066 N/mm², about 0.66 bar.

  • Uniformity: F_max / F_avg = 2.6 / 2.1 = 1.24. A ratio well above 1, or a glide curve that climbs along the stroke, flags uneven siliconization even when every value passes.

Use the agreed inner diameter and windows for your syringe format. The figures above are illustrative and are not ISO 11040-4 limits.

What to report

  • Reference to ISO 11040-4, with the edition (2024) and annex used.

  • Syringe description: volume, format, glass type, lubrication, lot and number of samples.

  • Stopper, plunger rod, fill liquid and fill volume, and the stoppering method and age.

  • Test speed, evaluation windows, load cell and fixture, and laboratory conditions.

  • Individual results and summary statistics: break-loose force, maximum and mean glide force, or peak force for breakage, pull-out and pull-off tests.

  • Failure observations: where the flange or cone broke, needle bond failure location, and any leakage.

  • Comparison with the 2024 minimum or maximum values where the annex sets them, or with the agreed specification.

Common ISO 11040-4 testing errors

  • Off-axis loading. An unguided ram or a loose barrel seat adds side load, inflates glide force and can crack the flange early.

  • Wrong windows. Including the end-of-stroke rise in the glide window, or the break-loose peak, distorts F_max and F_avg.

  • Low sampling rate. Below about 500 Hz the break-loose peak is under-read.

  • Ignoring syringe age. Break-loose rises with time after stoppering. Compare like with like.

  • Using 2015 criteria. Acceptance limits copied from older reports ignore the minimum and maximum values added in 2024.

  • Mixing standards. Testing a finished, drug-filled product to ISO 11040-4 Annex E instead of ISO 11040-8, or applying ISO 7886-1 piston forces to a prefilled glass syringe.

  • Metal-to-glass contact. Hard seats chip glass and start cracks. PTFE-lined seats prevent this.

  • Unqualified equipment. The 2024 edition requires qualified equipment and validated methods, so keep calibration and IQ/OQ records.

ISO 11040-4 vs related ISO and ASTM standards

  • ISO 11040-8: requirements and test methods for finished prefilled syringes, including break-loose and extrusion force with the drug product. The 2026 revision added a dedicated annex for that test.

  • ISO 11040-6: plastic barrels for injectables and sterilized subassembled syringes ready for filling. Many of its test methods mirror Part 4.

  • ISO 11040-5: plunger stoppers for injectables. ISO 11040-7: packaging systems for sterilized subassembled syringes ready for filling.

  • ISO 7886-1: sterile hypodermic syringes for single use, for manual use. It includes forces to operate the piston for conventional plastic syringes, not prefilled glass syringes.

  • ISO 80369-7: small-bore Luer connectors for intravascular or hypodermic use, with dimensions and performance. ISO 11040-4 requires its Luer front ends to conform. Connector performance tests (leakage, separation under axial load, unscrewing torque, ease of assembly, resistance to overriding and stress cracking) are in ISO 80369-20.

  • ISO 7864 and ISO 9626: hypodermic needles and needle tubing, the source of needle bond and tubing requirements.

  • ASTM: there is no ASTM equivalent to ISO 11040-4. US pharmacopoeia chapters on elastomeric components reference ISO 11040-8 for functional testing.

For other methods, see our ISO standards hub, ASTM standards hub and compression testing and puncture testing application pages.

ISO 11040-4 equipment summary

  • Fixture: Universal Grip prefilled glass syringe fixture (UGC-ISO11040-4-PF), 500 N standard, PTFE inserts for 0.5–10 mL syringes. See compression fixtures and adapters.

  • QC frame: Mark-10 F105 for glide and break-loose force, or F305 with headroom for flange breakage.

  • Central lab frame: Galdabini 2.5 kN benchtop UTM with a low-capacity load cell.

  • Software: test software for windowed break-loose and glide evaluation.

  • Not needed: extensometers. Environmental chambers only by agreement.

ISO 11040-4 FAQ

What is ISO 11040-4?

ISO 11040-4 is the international standard for glass barrels for injectables and sterilized subassembled syringes ready for filling. It sets materials, dimensions, quality and performance requirements, and its annexes give mechanical test methods for the flange, front end, needle, Luer components and glide force.

What is the current edition of ISO 11040-4?

ISO 11040-4:2024, the fourth edition, published in June 2024. It replaces ISO 11040-4:2015 and adds minimum breakage forces, a maximum needle penetration force, staked-needle requirements and glide force testing for non-lubricated syringes.

Which annex covers glide force?

Annex E (informative) gives the glide force test method. Annex C covers barrel tests such as finger flange and front-end breakage, Annex F needle penetration, and Annex G front-end component tests such as needle pull-out, Luer lock collar and tip cap tests.

What is the difference between break-loose force and glide force?

Break-loose force is the initial peak needed to start the plunger stopper moving. Glide force is the force needed to keep it moving, reported as a maximum and a mean over an agreed glide window.

What test speed is used for glide force testing?

Published work against the 2015 annexes used 100 mm/min, and some programs agree on other speeds such as 50 mm/min. Confirm the speed in the 2024 annex and record it, because glide force depends on speed.

How many syringes should I test?

ISO 11040-4 does not fix a sample size for most tests. Published studies used 50 syringes per test, and at least 30 is recommended for breakage statistics. Lot release normally follows an agreed sampling plan.

Is ISO 11040-4 the same as ISO 11040-8?

No. Part 4 covers the glass barrel and the sterilized subassembled syringe before filling. Part 8 covers the finished prefilled syringe with drug product, including break-loose and extrusion force.

How does ISO 11040-4 relate to ISO 80369-7?

ISO 11040-4 requires Luer slip and Luer lock front ends to conform to ISO 80369-7 and ISO 80369-1. Connector performance test methods are in ISO 80369-20.

Is ISO 7886-1 used for prefilled syringes?

ISO 7886-1 covers sterile single-use hypodermic syringes for manual use, typically plastic syringes filled at the point of use. Prefilled glass syringes are tested to the ISO 11040 series instead.

Can a Mark-10 tester run ISO 11040-4 tests?

Yes. Glide and break-loose forces are a few newtons and the speeds used fall within the F-series range of 13–330 mm/min. An F105 suits glide testing, and an F305 adds headroom for flange breakage.

What does the Universal Grip syringe fixture do?

It holds a glass prefilled syringe vertically in an interchangeable PTFE insert with a lever clamp, while a guided ram applies axial force for break-loose and glide force testing. A clear chamber collects expelled liquid. It is rated 500 N as standard.

Do I need an extensometer for syringe testing?

No. Plunger travel is measured from crosshead displacement, which is all the break-loose and glide force evaluation needs.

Need a fixture, grips or a complete test system for ISO 11040-4? Contact Universal Grip with your material, specimen size and machine connection and we will quote the right setup.