Dental Implant Fluid Bath Loading Fixture – ISO 14801, 2 kN

Angle-adjustable stainless steel fixture with a liquid bath for static and fatigue loading of dental implants in tests to ISO 14801; max. load 2 kN.

Brand: Universal Grip  |  Model: UGC-LB-010  |  ● Configured to order

  • Max. load 2 kN (about 450 lbf) for static and dynamic (fatigue) loading

  • Test angle adjustable from 0 to 90° to set the inclination of the implant

  • Upper probe with a 20 mm (0.79 in) spherical seat; broader probe version on request

  • Liquid bath keeps the implant immersed in test fluid during the test

  • Stainless steel construction

  • Female clevis coupling for a Ø 15.9 mm (5/8 in) pin

  • Used for dental implant tests to ISO 14801 (DIN EN ISO 14801)

Standards: ISO 14801

Trusted by Global Companies

Angle-adjustable loading fixture with liquid bath for dental implant testing

The UGC-LB-010 is a loading fixture for static and dynamic (fatigue) tests on endosseous dental implants, used for tests to ISO 14801 (DIN EN ISO 14801). It clamps the implant at a set angle, loads it from above through a probe with a 20 mm (0.79 in) spherical seat, and surrounds it with a liquid bath so the test can run with the implant immersed in test fluid. The fixture is stainless steel and rated to 2 kN (about 450 lbf). It is intended for implant manufacturers, contract test labs and dental research groups that produce static strength and fatigue data on implant bodies, abutments and abutment screws.

 

How it works

The implant assembly is clamped in the fixture, which adjusts from 0 to 90° and holds the specimen at the chosen test angle. The crosshead or actuator of the testing machine then drives the upper probe onto the specimen. Because the angle is set on the fixture and not by packing or re-machining specimen holders, one fixture covers the inclined loading arrangement described in ISO 14801 as well as axial loading at 0° and other off-axis checks used in development work.

The liquid bath keeps the implant immersed for the whole test, which matters most in fatigue programs that have to run in fluid and can last for days. Whether the bath is heated is specified at order: the data sheet lists bath temperature as on request, so tell us if you need an unheated bath or a temperature-controlled one, and at what temperature.

 

Probes and versions

  • Standard probe: upper probe with a 20 mm spherical seat

  • Broad probe: a wider probe with a 20 mm spherical seat, available on request

The fixture is configured to order. We ask for the test angle range, the implant and abutment geometry and the bath requirement, so drawings of the implant system are the most useful thing to send with an inquiry.

 

Typical applications

  • Static loading of dental implant and abutment assemblies

  • Fatigue testing of implants, abutments and abutment screws

  • Design and material comparisons during implant development

  • Test data for medical device approval

 

Testing machine fit

The fixture connects through a female clevis for a Ø 15.9 mm (5/8 in) pin. For static tests, the 2 kN rating sits within the range of a Galdabini benchtop frame such as the 2.5 kN or 5 kN universal testing machine. Fatigue testing needs a machine built for cyclic loading at the frequency and cycle count of your program; tell us which machine you run and we will confirm the mounting before quoting. Other couplings and adapters are available on request.

For general tensile or compression tests in fluid at body temperature, see the UGC-LB-020 heated liquid bath, or browse all environmental chambers and liquid baths.

Send us the implant and abutment geometry, the test angle range, whether the bath should be heated and your machine coupling, and we will configure the fixture: request a quote.

Product video

At a glance

  • Max. load: 2 kN

  • Test type: Static and dynamic (fatigue) loading

  • Test angle: 0–90°, adjustable

  • Upper probe: Spherical seat, 20 mm

  • Environment: Liquid bath

  • Bath temperature: On request

Dental Implant Fluid Bath Loading Fixture Specifications

MAX. LOAD

2 kN

 

TEST TYPE

Static and dynamic (fatigue) loading

 

TEST ANGLE

0–90°, adjustable

 

UPPER PROBE

Spherical seat, 20 mm

 

ENVIRONMENT

Liquid bath

 

BATH TEMPERATURE

On request

 

COUPLING

Female clevis, Ø 15.9 mm (5/8 in) pin

 

MATERIAL

Stainless steel

 

STANDARD NAMED ON DATA SHEET

ISO 14801 (DIN EN ISO 14801)

 

PART REF.

UGC-LB-010

 

Models & Options

  • Standard probe — Upper probe with 20 mm spherical seat (Specify at order)

  • Broad probe — Wider upper probe with 20 mm spherical seat (On request)

  • Liquid bath — Heated or unheated; bath temperature specified with the inquiry (On request)

  • Test angle range — Adjustable 0–90°; state the angle range you need (Specify at order)

  • Coupling and adapters — Female clevis for Ø 15.9 mm (5/8 in) pin as listed; other couplings and adapters available (On request)

What's Included

  • Angle-adjustable stainless steel loading fixture for dental implants, 0–90°

  • Liquid bath that keeps the implant immersed during the test

  • Upper probe with 20 mm spherical seat (standard or broad version, specified at order)

  • Female clevis coupling for a Ø 15.9 mm (5/8 in) pin

  • Bath heating is not listed on the data sheet; specify heated or unheated at order

  • Dimensional drawings and 3D models available on request

Supported Test Standards

  • ISO 14801 — Dynamic loading test for endosseous dental implants

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-LB-020 Heated Liquid Bath 400 x 400 mm

UGC-LB-020 Heated Liquid Bath 400 x 400 mm

Transparent polycarbonate liquid bath heated to +37 °C that keeps specimens and grips immersed during tensile or compres

UGC-LB-040 Unheated Water Bath 400 x 200 mm

UGC-LB-040 Unheated Water Bath 400 x 200 mm

Transparent polycarbonate water bath for tensile or compression tests on immersed specimens at ambient temperature, with

UGC-LB-050 Immersion Container with Compression Platens

UGC-LB-050 Immersion Container with Compression Platens

Clear acrylic immersion container with 96 mm stainless steel compression platens for compression tests on specimens subm

UGC-LB-060 Immersion Container with Tensile Grips for Tendons

UGC-LB-060 Immersion Container with Tensile Grips for Tendons

Clear acrylic immersion container with stainless steel tensile grips for cyclic and creep loading tests on tendons and s

Dental Implant Fluid Bath Loading Fixture FAQ

Which standard is the UGC-LB-010 used for?

It is a loading fixture for static and dynamic tests on endosseous dental implants to ISO 14801 (DIN EN ISO 14801). That is the only standard named on the data sheet.

 

What is the maximum load?

2 kN (about 450 lbf), for static and dynamic (fatigue) loading.

 

What angle can the implant be tested at?

The test angle is adjustable from 0 to 90°. State the angle range you need when you order.

 

Is the liquid bath heated?

The data sheet lists bath temperature as on request. Tell us whether you need a heated or an unheated bath, and the target temperature, and we will quote it that way.

 

What is the difference between the standard and the broad probe?

Both have a 20 mm spherical seat; the broad version is a wider probe. It is available on request.

 

Which testing machine does it fit?

The coupling is a female clevis for a Ø 15.9 mm (5/8 in) pin, and adapters are available. For static tests a Galdabini 2.5 kN or 5 kN benchtop frame covers the 2 kN rating; fatigue tests need a machine built for cyclic loading, and we confirm the mounting at quote.

 

What do you need from me to quote?

The implant and abutment geometry, the test angle range, the bath requirement (heated or unheated) and your machine coupling. Contact us with those details or call 978-882-1480.

 

Need help configuring a liquid bath?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.