Home › Products › Standard-Specific Test Fixtures › Dental Restorative Bond Shear Fixture





Dental Restorative Bond Shear Fixture – 5 kN
Stainless steel shear fixture, rated to 5 kN, for bond strength between dental restorative materials and tooth structure, with two shear inserts.
Brand: Universal Grip | Model: UGC-ST-190 | ● Configured to order
Designed for dental shear bond strength tests similar to ISO 29022
Notched shear insert with R1.6 mm cutting edge supplied
Flat shear insert with 2 x 10 mm cutting edge supplied
Rated for loads up to 5 kN (about 1,124 lbf)
Stainless steel construction for wet and clinical specimens
Female clevis coupling for a Ø 20 mm pin mounts directly on the machine
Standards: ISO 29022
Trusted by Global Companies


























Shear bond fixture for dental restorative materials on tooth structure
The UGC-ST-190 is a shear test fixture for determining the adhesive bond strength between direct dental restorative materials and tooth structure. It is used by dental adhesive developers, restorative material manufacturers and dental school research groups that bond a cylinder of restorative to a tooth surface and then shear it off under a measured load. The arrangement is similar to ISO 29022. The fixture is rated to 5 kN (about 1,124 lbf) and is made from stainless steel.
How it works
The bonded specimen is held in the fixture while a shear insert, driven by the testing machine, loads the restorative cylinder at the bond interface. Loading at the interface keeps the force acting on the bond itself, and the machine records the load at which the cylinder separates from the tooth structure.
Two shear inserts supplied
Notched insert, R1.6 mm: a cutting edge with a 1.6 mm (0.063 in) radius notch for loading a round restorative cylinder.
Flat insert, 2 x 10 mm: a straight cutting edge measuring 2 x 10 mm (about 0.08 x 0.39 in) for flat-edge shear loading.
Both inserts come with the fixture, so a lab can run a notched-edge test and a flat-edge comparison on the same hardware without buying a second fixture.
Stainless steel for wet specimens
The fixture is made from stainless steel, which suits wet and clinical specimens and is easy to wipe down between tests. A temperature range is not published for this product; if you plan to test above or below room temperature, tell us and we will confirm it at quote.
Machine fit
The UGC-ST-190 mounts directly with a female clevis for a Ø 20 mm pin. Other machine connections are covered by our adapters. With its 5 kN rating the fixture fits a Galdabini 5 kN benchtop universal testing machine or a Mark-10 F-series motorized frame such as the 1,500 lb tensile tester. Bond failure loads are often well below the fixture rating, so size the load cell for the forces you expect and not for the 5 kN limit.
Typical applications
Dental adhesive development and comparison of bonding systems
Restorative material testing
Dental school research
Bond strength QC on production lots
For bond tests on denture teeth, see the UGC-ST-830 denture tooth bond fixture; our general-purpose shear fixtures cover other materials.
Send us your specimen mounting dimensions and machine coupling, and we will confirm how the specimen is held and which adapter, if any, you need. Request a quote.
Dental Restorative Bond Shear Fixture Specifications
MAX. LOAD
5 kN (approx. 1,124 lbf)
TEST TYPE
Shear bond strength
STANDARD
Similar to ISO 29022
SPECIMEN
Direct dental restorative material bonded to tooth structure
SHEAR INSERTS
Notched R1.6 mm; flat 2 x 10 mm
COUPLING
Female clevis, Ø 20 mm pin
MATERIAL
Stainless steel
TEMPERATURE RANGE
On request
DRAWINGS
Dimensional drawings and 3D models on request
PART REF.
UGC-ST-190
Models & Options
Notched shear insert — R1.6 mm notched cutting edge; supplied with the fixture (Specify at order)
Flat shear insert — 2 x 10 mm flat cutting edge; supplied with the fixture (Specify at order)
Specimen mounting — Specimen holding matched to the mounting dimensions you specify (On request)
Machine coupling — Other couplings and adapters in place of the Ø 20 mm clevis (On request)
What's Included
Shear bond fixture in stainless steel
Notched shear insert with R1.6 mm cutting edge
Flat shear insert with 2 x 10 mm cutting edge
Female clevis coupling for a Ø 20 mm pin
Dimensional drawings or 3D models available on request
Supported Test Standards
ISO 29022 — Dental adhesion shear bond strength test; this fixture is similar to that arrangement
Compatible Grips & Fixtures
5 kN Benchtop Universal Testing Machine — Galdabini frame matching the 5 kN fixture rating
1,500 lb Tensile Tester — Mark-10 F-series motorized frame
Adapters — Couplings for machines without a Ø 20 mm clevis
UGC-ST-830 Denture Tooth Bond Fixture — Bond tests on polymer denture teeth
Shear Fixtures — Shear fixtures for other materials
Adhesives Testing — Fixtures for adhesive bond tests
“Customer quote”
Related Equipment
UGC-ST-830 Denture Tooth Bond Fixture ISO TS 19736
Angle-adjustable fixture, rated to 2 kN, for bond strength tests between polymer denture teeth and denture base material
UGC-ST-180 Toothbrush Test Fixture ISO 8627
Tuft stiffness fixture for manual toothbrushes with a micrometer-positioned brush clamp, guided loading column and dial
UGC-ST-200 Hip Prosthesis Endurance Fixture ISO 7206-4
Endurance test fixture, rated to 30 kN, for stemmed femoral components of hip prostheses, with pot holder, angle-adjusta
UGC-ST-170 Dental Drill and Turbine Test Fixture ISO 8325
Nickel-plated steel fixture with drill chuck and flat-tip punch for load tests on rotary dental instruments, burs and tu
Dental Restorative Bond Shear Fixture FAQ
What is the UGC-ST-190 used for?
It determines the shear bond strength between direct dental restorative materials and tooth structure. The bonded specimen is held while a shear insert loads the restorative cylinder at the bond interface.
Is it an ISO 29022 fixture?
The arrangement is similar to ISO 29022. If your protocol needs a specific geometry, send it to us and we will check it against the fixture before you order.
Which shear inserts come with it?
Two: a notched insert with an R1.6 mm cutting edge and a flat insert with a 2 x 10 mm cutting edge.
What is the maximum load?
5 kN (about 1,124 lbf). Bond failure loads are often much lower, so choose a load cell range that suits your expected forces.
Can it be used with wet specimens?
Yes. The fixture is stainless steel, which suits wet and clinical specimens. The temperature range is on request.
How does it mount on my testing machine?
With a female clevis for a Ø 20 mm pin. Other couplings are handled with our adapters; contact us with your machine model.