Home › Products › Standard-Specific Test Fixtures › Hypodermic Needle Tensile Fixture ISO 7864





Hypodermic Needle Tensile Fixture ISO 7864
Drill-chuck tensile fixture on a cardan joint for needle tube-to-hub bond tests on single-use hypodermic needles, rated to 2 kN.
Brand: Universal Grip | Model: UGC-ST-110 | ● Configured to order
Designed for needle tube-to-hub bond tests such as ISO 7864
Quick-action drill chuck clamps needle tubes from 0 to 6.5 mm (0 to 0.26 in)
Cardan (universal) joint lines the needle up with the load axis
Rated for tensile loads up to 2 kN (about 450 lbf)
Aluminum and steel construction keeps the fixture light
Female clevis coupling for a Ø 15.9 mm (5/8 in) pin; other couplings via adapters
Standards: ISO 7864
Trusted by Global Companies


























Drill-chuck tensile fixture for hypodermic needle hub bond tests
The UGC-ST-110 is a tensile test fixture for single-use sterile hypodermic needles. It is designed for the needle tube-to-hub bond test in ISO 7864: the needle tube is gripped and pulled in line with the hub, and the testing machine records the force at which the bond between tube and hub gives way. Needle manufacturers and medical device QC labs use it for routine bond strength checks. The fixture is rated for tensile loads up to 2 kN (about 450 lbf).
How it works
The upper part is a quick-action drill chuck with a clamping range of 0 to 6.5 mm (0 to 0.26 in). The chuck closes concentrically, so the needle tube is gripped on center whatever its diameter, and there are no jaw faces to swap when you change needle gauge. The quick-action chuck also keeps specimen changes short.
The chuck hangs from a cardan (universal) joint. The joint lets the chuck swing freely and settle on the load axis as force builds, so the tube is pulled straight out of the hub and side loads on a thin, easily bent specimen are kept out of the result.
The hub stays on the lower side of the load train while the chuck pulls the tube. How the hub is supported depends on the hub type, so we ask for needle gauge and hub type with every inquiry and confirm the lower holding arrangement in the quote.
Construction and machine fit
Aluminum and steel parts keep the fixture light, which helps when it hangs from a low-capacity load cell. It mounts through a female clevis with a Ø 15.9 mm (5/8 in) pin, and other machine connections are covered by our adapters. The temperature range is on request.
The 2 kN rating matches low-force frames such as a Mark-10 F-series motorized 500 lb tensile tester or a Galdabini 2.5 kN benchtop universal testing machine.
Typical applications
Hypodermic needle QC
Needle hub bond testing
Medical device manufacturing
Test programs that support regulatory submissions
Related needle fixtures
Bond strength is one of several mechanical checks on a needle. For 90° bend tests on finished needles see the UGC-ST-150 syringe needle test fixture, and for stiffness tests on the tubing itself see the UGC-ST-130 needle tubing bend fixture. The wider range is on our standard-specific test fixtures page.
Send us your needle gauges, hub type and machine coupling and we will confirm the configuration. Request a quote.
Hypodermic Needle Tensile Fixture ISO 7864 Specifications
MAX. LOAD
2 kN (approx. 450 lbf)
TEST TYPE
Tensile, needle tube-to-hub bond
STANDARD
ISO 7864
UPPER CLAMP
Quick-action drill chuck, 0–6.5 mm (0–0.26 in)
ALIGNMENT
Cardan (universal) joint
COUPLING
Female clevis, Ø 15.9 mm (5/8 in) pin
MATERIAL
Aluminum, steel
TEMPERATURE RANGE
On request
DRAWINGS
Dimensional drawings and 3D models on request
PART REF.
UGC-ST-110
Models & Options
Needle tensile fixture — Upper drill chuck 0–6.5 mm on cardan joint with clevis coupling (Specify at order)
Hub holding — Lower holding arrangement confirmed for your needle gauge and hub type (On request)
Machine coupling — Other couplings and adapters in place of the Ø 15.9 mm clevis (On request)
What's Included
Upper part with quick-action drill chuck, clamping range 0–6.5 mm
Cardan (universal) joint for alignment
Female clevis coupling for a Ø 15.9 mm (5/8 in) pin
Lower hub holding is confirmed at quote for your hub type
Dimensional drawings or 3D models available on request
Supported Test Standards
ISO 7864 — Sterile hypodermic needles for single use
Compatible Grips & Fixtures
UGC-ST-150 Syringe Needle Test Fixture — 90° bend tests on needles held by the hub
UGC-ST-130 Needle Tubing Bend Fixture — Stiffness tests on stainless steel needle tubing
500 lb Tensile Tester — Mark-10 F-series motorized frame in the same force range
2.5 kN Benchtop Universal Testing Machine — Galdabini benchtop frame for low-force tensile tests
Adapters — Couplings for other machine connections
“Customer quote”
Related Equipment
UGC-ST-130 Needle Tubing Bend Fixture
Bend test fixture for stiffness tests on stainless steel needle tubing, capillary tubes and cannulas, rated to 1 kN.
UGC-ST-150 Syringe Needle Test Fixture
Stainless steel fixture with spring-loaded hub holder and fork-shaped loading member for 90° bend tests on syringe needl
UGC-ST-100 Syringe Tensile Holder
Anodized aluminum upper holder that supports a syringe by its finger flange for plunger pull-out and flange strength tes
UGC-ST-120 Syringe Cartridge Test Fixture ISO 11608-3
Upper plunger rod and lower cartridge holder for stopper force tests on pen-injector cartridges, rated to 100 N.
Hypodermic Needle Tensile Fixture ISO 7864 FAQ
What does the UGC-ST-110 measure?
The bond strength between the needle tube and the hub of single-use hypodermic needles. The tube is pulled in line with the hub until the bond fails.
Which standard is it used with?
It is designed for needle tube-to-hub bond tests such as ISO 7864 for sterile hypodermic needles for single use.
What needle diameters can the chuck hold?
The quick-action drill chuck clamps from 0 to 6.5 mm (0 to 0.26 in), so no jaw change is needed between needle gauges.
Why is the chuck mounted on a cardan joint?
The joint lets the chuck settle on the load axis, so the needle tube is pulled straight and bending is kept out of the measurement.
How is the hub held?
The hub is held on the lower side of the load train. Tell us your needle gauge and hub type and we will confirm the arrangement with the quote.
What is the load rating and coupling?
2 kN (about 450 lbf), with a female clevis for a Ø 15.9 mm (5/8 in) pin. Other machine connections are available through adapters.