top of page

Home  ›  Products  ›  Standard-Specific Test Fixtures  ›  Syringe Cartridge Test Fixture ISO 11608-3

UGC-ST-120 syringe cartridge test fixture with plunger rod and cartridge holder by Universal Grip
UGC-ST-120 syringe cartridge test fixture with plunger rod and cartridge holder by Universal Grip
UGC-ST-120 Syringe Cartridge Test Fixture ISO 11608-3 by Universal Grip, view 2
UGC-ST-120 Syringe Cartridge Test Fixture ISO 11608-3 by Universal Grip, view 3
Product photo 4

Syringe Cartridge Test Fixture ISO 11608-3

Upper plunger rod and lower cartridge holder for stopper force tests on pen-injector cartridges, rated to 100 N.

Brand: Universal Grip  |  Model: UGC-ST-120  |  ● Configured to order

  • Designed for cartridge tests such as ISO 11608-3 on needle-based injection systems

  • Upper plunger rod acts directly on the cartridge stopper in compression

  • Lower holder with lined clamping jaws secures the cartridge on a base

  • Rated for 100 N (about 22.5 lbf) for sensitive low-force measurements

  • Nickel-plated steel construction for durability and corrosion resistance

  • Female clevis coupling for a Ø 15.9 mm (5/8 in) pin; other couplings via adapters

Standards: ISO 11608-3

Trusted by Global Companies

Plunger and cartridge holder for pen-injector cartridge tests

The UGC-ST-120 is a two-part compression fixture for the cartridges used in needle-based injection systems such as pen injectors. It is designed for cartridge tests in ISO 11608-3 and is used by pharmaceutical primary packaging labs, drug delivery developers and medical device QC teams who need to measure the force it takes to move the cartridge stopper. It is supplied as a matched pair, the plunger and the cartridge holder. The fixture is rated for 100 N (about 22.5 lbf), in keeping with the small forces involved.

 

How it works

The lower part is a cartridge holder mounted on a base. Its lined clamping jaws close on the cartridge and secure it under the load axis. The upper part is a plunger rod that travels with the crosshead and acts directly on the cartridge stopper.

As the testing machine advances, the plunger pushes the stopper along the cartridge in compression while force and travel are recorded. Forces such as stopper break-loose and glide force can then be read from the curve. Because the cartridge is clamped in the holder instead of standing loose, it stays in position under the plunger for the whole stroke.

 

Low-force construction

Both parts are nickel-plated steel for durability and corrosion resistance. Because the rating is 100 N, pair the fixture with a low-capacity load cell so that small stopper forces are resolved clearly. The temperature range is on request.

 

Mounting and machine fit

The fixture connects through a female clevis with a Ø 15.9 mm (5/8 in) pin. Other couplings are available through our adapters. In this force range it suits a Mark-10 F-series motorized 100 lb tester or a Galdabini 2.5 kN benchtop universal testing machine fitted with a small load cell.

 

Typical applications

  • Pen-injector cartridges

  • Pharmaceutical primary packaging

  • Medical device QC

  • Drug delivery development

 

What to specify

The plunger and holder are matched to the cartridge, so send the cartridge diameter and length along with your machine coupling. Custom versions for other cartridge formats and other couplings are available on request, as are dimensional drawings and 3D models. If you test single-use hypodermic syringes instead of cartridges, see the UGC-ST-160 syringe holder with interchangeable inserts. Other single-purpose fixtures are on our standard-specific test fixtures page, and general-purpose tooling is under compression fixtures.

Give us your cartridge dimensions and machine model and we will quote the plunger and cartridge holder as a set. Request a quote.

Product video

At a glance

  • Max. load: 100 N (approx. 22.5 lbf)

  • Test type: Compression, stopper force

  • Standard: ISO 11608-3

  • Scope of supply: Plunger and cartridge holder

  • Specimen holding: Lower holder with lined clamping jaws on a base

  • Coupling: Female clevis, Ø 15.9 mm (5/8 in) pin

Syringe Cartridge Test Fixture ISO 11608-3 Specifications

MAX. LOAD

100 N (approx. 22.5 lbf)

 

TEST TYPE

Compression, stopper force

 

STANDARD

ISO 11608-3

 

SCOPE OF SUPPLY

Plunger and cartridge holder

 

SPECIMEN HOLDING

Lower holder with lined clamping jaws on a base

 

COUPLING

Female clevis, Ø 15.9 mm (5/8 in) pin

 

MATERIAL

Steel, nickel-plated

 

TEMPERATURE RANGE

On request

 

DRAWINGS

Dimensional drawings and 3D models on request

 

PART REF.

UGC-ST-120

 

Models & Options

  • Cartridge test fixture — Upper plunger rod and lower cartridge holder, made for your cartridge diameter and length (Specify at order)

  • Machine coupling — Other couplings and adapters in place of the Ø 15.9 mm clevis (On request)

  • Custom cartridge sizes — Plunger and holder adapted to other cartridge formats (On request)

What's Included

  • Upper plunger rod

  • Lower cartridge holder with lined clamping jaws on a base

  • Female clevis coupling for a Ø 15.9 mm (5/8 in) pin

  • Dimensional drawings or 3D models available on request

Supported Test Standards

  • ISO 11608-3 — Needle-based injection systems for medical use – cartridges and containers

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-ST-160 Syringe Holder with Interchangeable Inserts

UGC-ST-160 Syringe Holder with Interchangeable Inserts

Nickel-plated steel syringe holder with interchangeable inserts for tension and compression tests on single-use syringes

UGC-ST-110 Hypodermic Needle Tensile Fixture ISO 7864

UGC-ST-110 Hypodermic Needle Tensile Fixture ISO 7864

Drill-chuck tensile fixture on a cardan joint for needle tube-to-hub bond tests on single-use hypodermic needles, rated

UGC-ST-130 Needle Tubing Bend Fixture

UGC-ST-130 Needle Tubing Bend Fixture

Bend test fixture for stiffness tests on stainless steel needle tubing, capillary tubes and cannulas, rated to 1 kN.

UGC-ST-100 Syringe Tensile Holder

UGC-ST-100 Syringe Tensile Holder

Anodized aluminum upper holder that supports a syringe by its finger flange for plunger pull-out and flange strength tes

Syringe Cartridge Test Fixture ISO 11608-3 FAQ

What is the UGC-ST-120 used for?

Compression tests on the cartridges of needle-based injection systems such as pen injectors. The plunger moves the cartridge stopper so that forces such as break-loose and glide force can be measured.

 

Which standard is it designed for?

Cartridge tests such as ISO 11608-3 for needle-based injection systems.

 

What does the fixture consist of?

An upper plunger rod and a lower cartridge holder with lined clamping jaws on a base. Both parts are nickel-plated steel.

 

What is the maximum load?

100 N (about 22.5 lbf). Use a low-capacity load cell so small stopper forces read clearly.

 

Which cartridge sizes can be tested?

The fixture is made for the cartridge you specify. Send the cartridge diameter and length with your inquiry.

 

Which testing machines does it fit?

It mounts with a female clevis and Ø 15.9 mm (5/8 in) pin and suits low-force frames such as our 100 lb tester. Other couplings are available through adapters.

 

Need help configuring a standard-specific test fixture?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.

bottom of page