Home › Products › Standard-Specific Test Fixtures › Wire Insulation Notch Resistance Fixture





Wire Insulation Notch Resistance Fixture – Cut-Through, 2 kN
Cut-through fixture for notch resistance testing of wire and cable insulation, designed for LV 112 and ASTM D3032 tests; nickel-plated steel, 2 kN.
Brand: Universal Grip | Model: UGC-ST-610 | ● Configured to order
Defined cutting edge geometry per LV 112 Fig. 5 and ASTM D3032 Fig. 18
Lower holder positions the wire square to the cutting edge
Nickel-plated steel body and edge for corrosion resistance
Rated to 2 kN (450 lbf) in compression for thick insulation layers
Optional cutting edge version for hook-up wire per ASTM D3032 Fig. 19
EN 50618 version with 0.45 mm radius fin for cable insulated on both sides, 1 kN
Female clevis coupling with Ø 15.9 mm (5/8 in) pin
Standards: LV 112 · ASTM D3032 · EN 50618
Trusted by Global Companies


























Cut-through fixture for notch resistance of wire insulation
The UGC-ST-610 measures the notch resistance of insulated wire: the force a defined cutting edge needs to break through the insulation. It is designed for cut-through tests such as LV 112 and ASTM D3032 and serves automotive wiring harness suppliers, hook-up wire qualification labs, photovoltaic cable producers and cable manufacturing QC.
How it works
The wire specimen lies on a lower holder that positions it square to the cutting edge. The operator loads the specimen by hand. The testing machine then drives the edge into the insulation in compression until the edge breaks through. Because the edge geometry is defined by the figure in the specification, results from different labs and batches can be compared.
Versions
LV 112 Fig. 5 / ASTM D3032 Fig. 18 – cut-through fixture for wire insulation, 2 kN (450 lbf)
ASTM D3032 Fig. 19 – hook-up wire version with optional cutting edge, 2 kN (450 lbf)
EN 50618 version – upper fin with a 0.45 mm (0.018 in) radius edge and a lower holder, 1 kN (225 lbf), for cable insulated on both sides
The versions differ in cutting edge and rating, so name the standard and figure when you order, along with your wire diameter range. Dimensional drawings and 3D models are available on request, and custom versions can be quoted for other edge geometries.
Construction and ratings
Max. load: 2 kN (450 lbf); 1 kN for the EN 50618 version
Material: nickel-plated steel body and edge
Coupling: female clevis with Ø 15.9 mm (5/8 in) pin
Operation: manual specimen loading, machine-driven edge
Temperature range: on request
Typical applications
Automotive wiring harnesses
Hook-up wire qualification
Photovoltaic cable insulation
Cable manufacturing quality control
Machine fit
The fixture works in compression at up to 2 kN, which a Mark-10 500 lbF motorized tester or a Galdabini 2.5 kN universal testing machine covers. The clevis coupling takes a Ø 15.9 mm pin, and adapters connect it to other load strings. The data sheet does not list a device for detecting the moment of breakthrough, so tell us how your method detects it and we will confirm what the setup needs.
Related cable fixtures
For bending stiffness of complete automotive cables, see the UGC-ST-060 cable flexibility fixture. Other product-specific tools are listed under standard-specific test fixtures.
Tell us the standard and figure, your wire diameter range and your machine coupling and we'll configure the fixture — request a quote.
Wire Insulation Notch Resistance Fixture Specifications
MAX. LOAD
2 kN (450 lbf)
TEST TYPE
Cut-through (notch resistance), compression
SPECIMEN
Insulated wire and cable
CUTTING EDGE
Per LV 112 Fig. 5 / ASTM D3032 Fig. 18
MATERIAL
Steel, nickel-plated
COUPLING
Female clevis, Ø 15.9 mm (5/8 in) pin
TEMPERATURE RANGE
On request
OPERATION
Manual specimen loading, machine-driven edge
STANDARDS
LV 112, ASTM D3032
PART REF.
UGC-ST-610
Models & Options
LV 112 / ASTM D3032 Fig. 18 — Cut-through fixture for wire insulation, 2 kN (Specify at order)
ASTM D3032 Fig. 19 — Hook-up wire version with optional cutting edge, 2 kN (Specify at order)
EN 50618 version — Upper fin with 0.45 mm radius edge and lower holder, 1 kN, for cable insulated on both sides (Specify at order)
What's Included
Cut-through fixture with cutting edge for the standard and figure ordered
Lower holder for the wire specimen
Female clevis coupling for Ø 15.9 mm (5/8 in) pin
EN 50618 version: 1 upper fin and 1 lower holder
Supported Test Standards
LV 112 — Automotive wiring – notch resistance of insulation (Fig. 5)
ASTM D3032 — Hook-up wire insulation test methods (Figs. 18 and 19)
EN 50618 — Electric cables for photovoltaic systems
Compatible Grips & Fixtures
500 lb motorized tester — Mark-10 F-series tester covering the 2 kN rating
2.5 kN universal testing machine — Galdabini benchtop frame for compression at up to 2 kN
Adapters — Connect the Ø 15.9 mm clevis to other machine couplings
UGC-ST-060 cable flexibility fixture — Three-pulley bending fixture for automotive cables
Compression testing — Machines and fixtures for compression tests
“Customer quote”
Related Equipment
UGC-ST-060 Automotive Cable Flexibility Fixture ISO 19642-2
Three-pulley bending fixture for measuring the flexibility of automotive cables, designed for ISO 19642-2 tests, with pu
UGC-ST-590 Electrical Varnish Helical Coil Bond Fixture
Helical coil bond strength fixture for electrical insulating varnishes, designed for ASTM D2519 tests, with roller suppo
UGC-ST-270 Rebar Stress Corrosion Cracking Fixture GOST 34028
Free-standing dead-weight rig that holds reinforcing bar under constant tensile load in a 100 °C calcium ammonium nitrat
UGC-ST-090 Battery Handle Strength Fixture EN 50342-5
Pull test fixture for the carrying handles of lead-acid starter batteries, designed for EN 50342-5 clause 4.4.4 tests an
Wire Insulation Notch Resistance Fixture FAQ
What does the notch resistance fixture measure?
It measures the force needed for a defined cutting edge to break through the insulation of a wire or cable. The testing machine drives the edge into the insulation in compression.
Which specifications does it cover?
The standard version has a cutting edge per LV 112 Fig. 5 and ASTM D3032 Fig. 18. A hook-up wire version per ASTM D3032 Fig. 19 and an EN 50618 version are also available.
What is the maximum load?
2 kN (450 lbf) for the LV 112 and ASTM D3032 versions, and 1 kN (225 lbf) for the EN 50618 version.
What is different about the EN 50618 version?
It uses an upper fin with a 0.45 mm (0.018 in) radius edge and a lower holder, and is intended for cable insulated on both sides.
How is the wire held?
The wire lies on a lower holder that positions it square to the cutting edge. Specimens are loaded by hand.
Which testing machine suits it?
Any frame that runs in compression at up to 2 kN, such as a Mark-10 500 lbF tester or a Galdabini 2.5 kN universal testing machine. Adapters cover other couplings.
What should I specify when ordering?
The standard and figure (LV 112 Fig. 5, ASTM D3032 Fig. 18 or 19, or EN 50618), your wire diameter range and the machine coupling.