Gauge Length Selection for Video Extensometers Across ASTM Standards

  • christopherthomasb
  • 5 days ago
  • 5 min read

How to Select the proper Gauge Length for your tensile test


Introduction: Why Video Extensometer Gauge Length Matters in Optical Strain Measurement

In tensile and mechanical testing, strain measurement accuracy is only as good as the gauge length definition used to calculate it. Gauge length directly determines how deformation is interpreted, how localized strain is captured, and whether test results comply with ASTM and ISO standards. While traditional clip-on extensometers rely on fixed mechanical gauge lengths, video extensometers introduce a fundamentally different and more flexible approach through non-contact, optical measurement.


Video extensometers allow users to define virtual gauge lengths, dynamically track deformation, and measure strain across multiple regions of a specimen simultaneously. These capabilities make them especially valuable for testing plastics, rubbers, metals, and composites under ASTM standards where deformation behavior varies significantly. Selecting the correct gauge length is therefore a critical technical decision, not a trivial setup step.


This article provides a detailed, standards-driven guide to gauge length selection for video extensometers, with practical examples from ASTM D638, ASTM D412, ASTM E8, and ASTM D3039.

Video Extensometers vs. Clip-On Extensometers - Which is Best?

Fundamentals of Gauge Length in Video Extensometry

What Is a Virtual Gauge Length?

A virtual gauge length is the optically defined distance between two or more tracked points on a specimen surface. Unlike mechanical extensometers, video extensometers do not physically contact the specimen. Instead, high-resolution cameras track surface features, applied markers, or speckle patterns to calculate displacement and strain.

Key characteristics of virtual gauge lengths include:

  • Defined in software rather than hardware

  • Adjustable before or during testing

  • Not limited to standardized clip-on lengths

  • Immune to mass loading or specimen interference

This approach enables accurate strain measurement even in applications involving large elongation, fragile specimens, or aggressive failure modes.


Sources of Error Related to Improper Gauge Length Selection

Improper gauge length selection can significantly distort test results. Common issues include:

  • Grip influence contaminating strain data

  • Localized necking dominating average strain values

  • Edge effects near specimen shoulders

  • Over- or under-resolution of elastic strain

Video extensometers reduce—but do not eliminate—these risks. Correct gauge placement remains essential for ASTM compliance and data repeatability.


Fixed vs. Dynamic Gauge Lengths

Fixed Gauge Lengths

A fixed gauge length remains constant throughout the test and mirrors the behavior of a traditional extensometer, with the added benefits of non-contact measurement.

Advantages:

  • Direct alignment with ASTM-specified gauge lengths

  • Excellent repeatability for quality control environments

  • Simplified reporting and auditing

Limitations:

  • Cannot adapt to shifting deformation zones

  • May under-represent localized strain in necking materials

Fixed gauge lengths are commonly used for metals and standardized plastic testing, where deformation is relatively uniform up to yield.

Dynamic (Adaptive) Gauge Lengths

Dynamic gauge lengths allow tracking points to move or expand as the specimen elongates. This capability is particularly valuable for materials exhibiting non-uniform strain, such as elastomers or polymers with pronounced necking.

Benefits include:

  • Accurate tracking through large elongation

  • Improved failure and necking analysis

  • Reduced sensitivity to initial marker placement

Dynamic gauge lengths are frequently employed in ASTM D412 rubber testing and advanced R&D applications.

Video Extensometer Sample Marking Guide

Multi-Region and Multi-Gauge Strain Measurement

Why Multi-Region Measurement Matters

One of the most powerful advantages of video extensometers is the ability to define multiple simultaneous gauge lengths on a single specimen. This enables:

  • Detection of strain gradients

  • Comparison of elastic and plastic regions

  • Measurement of transverse strain for Poisson’s ratio

Typical Applications

  • Center vs. edge strain comparison

  • Longitudinal and transverse strain tracking

  • Failure mode identification in composites

Multi-region measurement is especially valuable in composites and advanced polymer testing, where deformation is rarely uniform.

Gauge Length Selection Across Common ASTM Standards

ASTM D638 Sample Size and Gauge Length Guide
ASTM D638 Sample Size and Gauge Length Guide

ASTM D638 specifies tensile testing for rigid and semi-rigid plastics using standardized dog-bone specimens. Gauge length selection must remain within the reduced section to avoid shoulder effects.

Key considerations:

  • Typical gauge lengths: 25–50 mm

  • Necking frequently occurs prior to fracture

  • Video extensometers avoid damage at break

  • Dynamic gauge lengths help track strain through necking

This makes video extensometers particularly effective for high-elongation thermoplastics.

ASTM D412 presents one of the most challenging strain measurement environments due to elongations exceeding 300–700%.

Challenges with contact extensometers include:

  • Slippage

  • Detachment at high strain

  • Damage to soft specimens

Video extensometers overcome these limitations by enabling:

  • Long virtual gauge lengths

  • Dynamic tracking throughout deformation

  • Multi-region analysis to identify tear initiation

For elastomers, dynamic gauge lengths are strongly preferred.

ASTM E8/E8M governs tensile testing of metallic materials and specifies standardized gauge lengths such as 50 mm or 2 inches.

Critical requirements include:

  • High strain resolution in the elastic region

  • Accurate yield point determination

  • Minimal noise and optical distortion

Fixed virtual gauge lengths aligned with ASTM E8 dimensions are typically used, provided the camera resolution and lighting support small elastic strain measurement.

ASTM D3039 addresses tensile testing of composite laminates, where materials exhibit anisotropic behavior.

Gauge length considerations:

  • Separate longitudinal and transverse strain regions

  • Dual-axis strain measurement

  • Multi-region gauges for failure analysis

Video extensometers excel here by enabling simultaneous axial and transverse strain tracking, supporting Poisson’s ratio calculations and advanced composite characterization.

Summary Table: Gauge Length Selection by ASTM Standard

ASTM Standard

Material Type

Typical Gauge Length

Preferred Mode

Key Notes

ASTM D638

Plastics

25–50 mm

Fixed / Dynamic

Necking common

ASTM D412

Rubber

25–100 mm

Dynamic

Large elongation

ASTM E8

Metals

50 mm / 2 in

Fixed

Elastic strain critical

ASTM D3039

Composites

25–50 mm

Multi-Region

Anisotropic behavior

Practical Guidelines for Gauge Length Selection

  • Place gauge lengths entirely within the reduced section

  • Avoid proximity to grips and shoulders

  • Match gauge length to expected strain range

  • Balance field of view with optical resolution

  • Verify marker contrast and lighting before testing

Proper setup dramatically improves repeatability, compliance, and data credibility.

Common Mistakes and How to Avoid Them

  • Gauge length too short for large elongation materials

  • Markers placed too close to grips

  • Insufficient resolution for low-strain metals

  • Ignoring transverse strain in composites

Best Practices for ASTM Compliance and Documentation

  • Record gauge length definition in test reports

  • Maintain consistency across test programs

  • Validate optical systems during method development

  • Ensure traceability during audits and certifications

Conclusion: Leveraging Video Extensometers for ASTM-Compliant Strain Measurement

Gauge length selection is a foundational element of accurate strain measurement, and video extensometers provide unmatched flexibility in addressing the diverse requirements of ASTM testing. Whether using fixed gauge lengths for metals, dynamic tracking for elastomers, or multi-region analysis for composites, optical extensometry enables higher data quality, broader test coverage, and improved laboratory efficiency.


By understanding how gauge length interacts with material behavior and ASTM standards, laboratories can fully leverage video extensometers as a precision strain measurement solution, not merely a convenience upgrade.


Visit our website www.universalgripco.com for more information. Or inquire here: https://www.universalgripco.com/contact