Home › Products › Video Extensometers › UGC-VE-4202 2 MPx Two-Unit Video Extensometer





UGC-VE-4202 2 MPx Two-Unit Video Extensometer
Two 2 MPx 400 mm VE units stacked and joined into one continuous field of view. A long field of view for plastics, film and composites.
Brand: Universal Grip | Model: UGC-VE-4202 | ● Configured to order
ISO 9513 Class 0.5 measuring length 2 × 110 mm; Class 1 2 × 220 mm
90 Hz typical sampling, up to 1 kHz with a narrowed view
All-in-one: camera, C-mount lens, blue LED light and USB light relay
Joined fields of view for long or high-elongation specimens
Non-contact: stays on through break, no knife edges
Live output to your test machine (digital, analog, pulse) and API control
Downloads: Video Extensometer Inquiry Form (Excel)
Standards: ASTM D638 · ASTM D882 · ASTM D3039 · ASTM D6272
Trusted by Global Companies


























Two 400 mm 2 MPx units stacked into one continuous field of view
The UGC-VE-4202 is a non-contact optical (video) extensometer from the X-Sight 4000 Series, supplied, configured and supported by Universal Grip. It tracks gauge marks on the specimen with two 2 MPx cameras and reports strain, total length, delta length and angle in real time, so it can stay on the specimen through break with no knife edges to slip or damage. Two 4000 Series units with a joined calibration grid nearly double the measuring length, for very long or highly elongating specimens such as elastomers, films and wire.
Measuring length, accuracy class and sampling rate
Per ISO 9513 the UGC-VE-4202 measures 2 × 110 mm at Class 0.5, 2 × 220 mm at Class 1 and 720 mm at Class 2. Class 0.5 is the tight class labs choose for modulus and proof-strength work; ISO 6892-1 calls for Class 1 or better for yield and proof strength and allows Class 2 for elongation after fracture.
Typical sampling is 90 Hz (43 Hz with the full image), and narrowing the view width raises it to 1 kHz. Higher pixel counts trade frame rate for field of view: at this resolution the camera favours a balance of field size and speed.
As a sizing check: the Class 2 measuring length of 720 mm is long enough to follow a gauge section of about 0.7 m, to measure several gauge lengths on one long specimen, or to watch more than one specimen at once. We confirm the usable field from your specimen, grips and expected elongation.
What the UGC-VE-4202 is best for
plastics and film with high elongation to ASTM D638 and ASTM D882
composite and laminate coupons to ASTM D3039
flexural tests on long spans to ASTM D6272 with the bend-line probe
two-zone measurements on long specimens
Two 400 mm units for very long fields of view
Two 4000 Series units, each with its own 400 mm LED, are stacked and joined with one calibration grid into a continuous field of view. Each unit weighs 2.5 kg. Use it when a single camera cannot cover the full elongation, for example on rubber, film, wire and long cables.
Choosing the lens and working distance
Measuring length, lens focal length and working distance are linked: choose two and the third follows. For the Class 0.5 field of view (2 × 110 mm high by 70 mm wide) the working distances are 16 mm lens 134 mm, 25 mm lens 237 mm, 35 mm lens 322 mm, 50 mm lens 430 mm (from the front cover). Longer working distances reduce LED intensity and can need longer shutter times, so tell us your frame and chamber geometry and we will pick the lens.
Connects to your testing machine
Strain is streamed live to the test-machine controller over DOLI Binary, MODBUS, HP Video, TCP/IP or RS232, as an analog signal through auxiliary D/A converters, or as an encoder-style quadrature pulse. The JSON and MRT APIs let your test software start and stop measurement, switch methods and set the gauge length. That is how we retrofit video extensometers to Instron, MTS, ZwickRoell, Shimadzu, Test Resources, ADMET and Galdabini frames; see retrofitting a video extensometer to an existing machine.
Compare the UGC-VE-4202 with similar models
UGC-VE-4201: 4000 Series two-unit, 1 MPx, Class 0.5 2 × 80 mm, 200 Hz typical
UGC-VE-4205: 4000 Series two-unit, 5 MPx, Class 0.5 2 × 130 mm, 175 Hz typical
UGC-VE-2102: 2000 Series single-camera, 2 MPx, Class 0.5 110 mm, 90 Hz typical
UGC-VE-2202: 2000 Series dual-camera, 2 MPx, Class 0.5 2 × 110 mm, 90 Hz typical
Not sure which model fits? Read how to choose an extensometer, gauge length selection for video extensometers, or compare all models on our video extensometer page.
Video Extensometer vs. Clip-On Extensometers
At a glance
Cameras: 2 × 2 MPx industrial cameras
Sampling rate: 90 Hz typical · 43 Hz at full view · up to 1 kHz with reduced view width
Measuring length (ISO 9513): Class 0.5: 2 × 110 mm · Class 1: 2 × 220 mm · Class 2: 720 mm
Field of view, Class 0.5 (H × W): 2 × 110 × 70 mm
Field of view, Class 1 (H × W): 2 × 220 × 120 mm
Field of view, Class 2 (H × W): 720 × 238 mm
UGC-VE-4202 2 MPx Two-Unit Video Extensometer Specifications
MODEL
UGC-VE-4202 · 4000 Series, two VE units
CAMERAS
2 × 2 MPx industrial cameras
SAMPLING RATE
90 Hz typical · 43 Hz at full view · up to 1 kHz with reduced view width
MEASURING LENGTH (ISO 9513)
Class 0.5: 2 × 110 mm · Class 1: 2 × 220 mm · Class 2: 720 mm
FIELD OF VIEW, CLASS 0.5 (H × W)
2 × 110 × 70 mm
FIELD OF VIEW, CLASS 1 (H × W)
2 × 220 × 120 mm
FIELD OF VIEW, CLASS 2 (H × W)
720 × 238 mm
WORKING DISTANCE
300–500 mm typical from front cover; extendable on request
LENS
C-mount; 12, 16, 25, 35 or 50 mm focal length selected per application
LIGHT
2 × 400 mm blue LED (465 nm, 330 lm, 16 W), USB-switched
PC INTERFACE
USB 3.0 per camera (2 total) + USB 2.0 for light relay; 3 m cables standard
DATA OUTPUT
DOLI Binary, MODBUS, HP Video, TCP/IP, RS232; analog via D/A; quadrature pulse; JSON and MRT API
HOUSING DIMENSIONS
424 × 187 × 80 mm (L × W × H) per unit
WEIGHT
2.5 kg per unit
POWER
24 V DC via RJ45 power cable (802.3af Mode B); 20 W per unit
MOUNTING
Tripod 1/4"-20 UNC (center) or 2 × M6 on a 165 mm vertical span
ENVIRONMENT
Indoor; 5–40 °C, 30–70 % RH
SOFTWARE
Windows 10/11 64-bit or Windows Server 2019/2022; perpetual license on USB key
PC REQUIREMENTS
2 GHz dual-core min. (4 GHz, 8+ cores rec.); 4 GB RAM min. (16 GB rec.); OpenGL 3.0
STANDARDS
ISO 9513 accuracy classes; used for ISO 6892-1, ASTM E8 and similar methods
Models & Options
C-mount lens — 12, 16, 25, 35 or 50 mm focal length chosen from your gauge length and working distance (Specify at order)
AX module (included) — Point, Line and Extreme Line probes for axial strain and elongation (AX)
Transverse module — Transverse line and bend line for r-value, Poisson's ratio and deflection (TR)
Torsion / crack module — Torsion line and crack-opening probe (TO / CR)
Rebar module — Rebar line for reinforcing-steel testing (ITT)
DIC / post-processing — DIC area strain mapping and re-analysis of recorded tests (DIC / PP)
High-speed, device input, color — High-speed recorder, external force/temperature/pressure input, color value distribution (HS / DI / LVD)
Accessories — Calibration grids, tripods, polarization kits, PCIe USB 3.0 card, pulse generator, D/A converters (On request)
Sample marking — Gauge-mark marking tools for fast, repeatable marks (Marking tool)
What's Included
2 × VE unit with lens, 400 mm LED & built-in USB relay
Joined calibration grid sized to the application
Cable harness, 24 V power supply
Software on installation USB + USB license key
Axial software module (perpetual license)
1 year software updates & technical support
Supported Test Standards
ASTM D638 — Tensile properties of plastics
ASTM D882 — Tensile properties of thin plastic sheeting
ASTM D3039 — Tensile properties of polymer matrix composites
ASTM D6272 — Flexural properties by four-point bending
Compatible Grips & Fixtures
Universal testing machines — Retrofit to Instron, MTS, ZwickRoell, Shimadzu, Test Resources, ADMET or Galdabini frames
Tensile grips — Wedge, vise, roller and pneumatic grips; the camera views the specimen between them
Environmental chambers — Measure through the chamber window with the right lighting
Marking tool — Repeatable gauge marks for video extensometers
Clip-on extensometers — Contact alternative for single-gauge-length tests
All extensometers — Clip-on, long-travel, video and 3D extensometers
“Customer quote”
Related Equipment
UGC-VE-4201 1 MPx Two-Unit
4000 Series two-unit, 1 MPx
UGC-VE-4205 5 MPx Two-Unit
4000 Series two-unit, 5 MPx
UGC-VE-2102 2 MPx Single-Camera
2000 Series single-camera, 2 MPx
UGC-VE-2202 2 MPx Dual-Camera
2000 Series dual-camera, 2 MPx
UGC-VE-4202 2 MPx Two-Unit Video Extensometer FAQ
What gauge length and field of view does the UGC-VE-4202 cover?
Per ISO 9513, Class 0.5: 2 × 110 mm; Class 1: 2 × 220 mm; Class 2: 720 mm. The exact field depends on the lens and working distance, which we pick from your specimen and frame.
Which ISO 9513 class do I need?
ISO 6892-1 calls for Class 1 or better for yield and proof strength and allows Class 2 for elongation after fracture. Many labs use Class 0.5 for modulus. Plastics methods such as ISO 527 set their own accuracy requirements.
How fast does the UGC-VE-4202 sample?
90 Hz typical, 43 Hz with the full image, and up to 1 kHz with a narrowed view width. Data streams to the test machine controller in real time.
Will the UGC-VE-4202 work with my existing testing machine?
Yes, in most cases. It outputs DOLI Binary, MODBUS, HP Video, TCP/IP, RS232, analog via D/A converters or a quadrature pulse, and the JSON/MRT APIs allow remote start/stop and gauge-length setting. Tell us the make and model of your frame and controller. See retrofitting a video extensometer.
How is the UGC-VE-4202 different from the UGC-VE-4102-D?
The UGC-VE-4202 uses two complete 400 mm units, each with its own LED, stacked and joined into one field of view. The -D version puts two cameras in a single housing. Two units reach further; one housing is quicker to mount.
What software comes with the UGC-VE-4202?
The axial (AX) module with Point, Line and Extreme Line probes, on a perpetual license stored on a USB key, with one year of updates and technical support. Optional modules add transverse strain (r-value, Poisson's ratio), torsion, crack opening, rebar, DIC mapping and post-processing.
How do I mark specimens for a video extensometer?
Use high-contrast dots or lines at the gauge length. Our marking tools and the sample marking guide cover the common methods.
How do I get a quote for the UGC-VE-4202?
Send your material, test standard, specimen size and gauge length, expected elongation, and your testing machine make/model. Use the Request a Quote button, call 978-882-1480, or download the video extensometer inquiry form.