top of page

Home  ›  Products  ›  Puncture Fixtures  ›  Dental Ceramic Biaxial Flexure Fixture ISO 6872

UGC-PF-040 piston-on-three-balls biaxial flexure fixture for ISO 6872 dental ceramics by Universal Grip
UGC-PF-040 piston-on-three-balls biaxial flexure fixture for ISO 6872 dental ceramics by Universal Grip
UGC-PF-040 Dental Ceramic Biaxial Flexure Fixture ISO 6872 by Universal Grip, view 2
Product photo 3
Product photo 4

Dental Ceramic Biaxial Flexure Fixture ISO 6872

Piston-on-three-balls fixture with tungsten carbide piston for biaxial flexural strength of dental ceramic discs to ISO 6872, rated 2.5 kN.

Brand: Universal Grip  |  Model: UGC-PF-040  |  ● Configured to order

  • Piston-on-three-balls arrangement per ISO 6872

  • Three hardened steel support balls, Ø 3.175 mm (0.125 in)

  • Tungsten carbide loading piston, Ø 1.4 mm (0.055 in)

  • Stainless steel 304 body rated 2.5 kN (562 lbf)

  • Usable from -70 to +280 °C (-94 to 536 °F) inside a chamber

  • Version for ceramic substrates to ASTM F394 available

Standards: ISO 6872  ·  ASTM F394

Trusted by Global Companies

Biaxial flexure fixture for dental ceramics

The UGC-PF-040 is a piston-on-three-balls fixture for measuring the flexural strength of dental ceramics used in all-ceramic and metal-ceramic restorations, according to ISO 6872. Dental material manufacturers, dental lab R&D groups and university research labs use biaxial flexure because it loads a thin disc away from its edges, so the result reflects the material rather than edge chips left from cutting or grinding. The same stainless steel fixture is also available set up for ceramic substrates to ASTM F394.

 

How it works

A disc specimen rests on three hardened steel balls of Ø 3.175 mm (0.125 in) arranged on a support circle. A tungsten carbide piston of Ø 1.4 mm (0.055 in) presses on the center of the disc from above until it fractures. Three-point support keeps the disc stable even if it is not perfectly flat, and the carbide piston holds its contact geometry over many tests on hard materials such as zirconia. The fracture load, disc thickness and support geometry are then used to calculate biaxial flexural strength.

 

Built for chamber testing

The body is stainless steel 1.4301 (304) and is rated 2.5 kN (562 lbf). It can be used from -70 to +280 °C (-94 to 536 °F) inside an environmental chamber, so strength can be measured on conditioned specimens without moving them to a different fixture. At 1.23 kg (2.7 lb) it is easy to place and center in a chamber. The standard coupling is a female clevis with a Ø 15.9 mm (5/8 in) pin; others are available on request or through adapters. The fixture is supplied with its three support balls.

 

Versions

  • ISO 6872: dental ceramics

  • ASTM F394: ceramic substrates, modulus of rupture

When ordering, specify the standard, your disc specimen diameter and the machine coupling.

 

Typical applications

  • Dental ceramics and zirconia

  • Restoration materials

  • Ceramic substrates

  • Dental lab R&D

 

Machine fit

With a 2.5 kN rating, the fixture suits a Galdabini 2.5 kN or 5 kN benchtop UTM, and its full capacity is also within the range of a Mark-10 F-series motorized tester (up to 1,500 lbF). For three- and four-point bending of bar specimens, see our bend testing fixtures.

Tell us the standard, disc diameter and machine coupling, and we will configure the fixture: request a quote.

Product video

At a glance

  • Max. load: 2.5 kN

  • Test type: Biaxial flexure (piston on three balls)

  • Support: 3 hardened steel balls, Ø 3.175 mm

  • Piston: Ø 1.4 mm, tungsten carbide

  • Coupling: Female clevis, Ø 15.9 mm (5/8 in) pin; others on request

  • Material: Stainless steel 1.4301 (304)

Dental Ceramic Biaxial Flexure Fixture ISO 6872 Specifications

MAX. LOAD

2.5 kN

 

TEST TYPE

Biaxial flexure (piston on three balls)

 

SUPPORT

3 hardened steel balls, Ø 3.175 mm

 

PISTON

Ø 1.4 mm, tungsten carbide

 

COUPLING

Female clevis, Ø 15.9 mm (5/8 in) pin; others on request

 

MATERIAL

Stainless steel 1.4301 (304)

 

TEMPERATURE RANGE

-70 to +280 °C

 

WEIGHT

1.23 kg

 

SCOPE OF SUPPLY

1 fixture incl. 3 balls

 

DRAWINGS

Dimensional drawing on data sheet; certified drawings and 3D models on request

 

PART REF.

UGC-PF-040

 

Models & Options

  • ISO 6872 version — Dental ceramics (Specify at order)

  • ASTM F394 version — Ceramic substrates, modulus of rupture (Specify at order)

  • Disc specimen diameter — Support set up for your disc size (Specify at order)

  • Machine coupling — Female clevis Ø 15.9 mm standard; others (On request)

What's Included

  • 1 biaxial flexure fixture in stainless steel

  • 3 hardened steel support balls, Ø 3.175 mm

  • Tungsten carbide loading piston, Ø 1.4 mm

Supported Test Standards

  • ISO 6872 — Dentistry, ceramic materials (biaxial flexural strength)

  • ASTM F394 — Biaxial flexure strength (modulus of rupture) of ceramic substrates

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-PF-030 Geomembrane Index Puncture Fixture

UGC-PF-030 Geomembrane Index Puncture Fixture

Ring clamp with 45 mm opening and hardened 8 mm probe for index puncture tests on geomembranes and geosynthetics to ASTM

UGC-PF-050 Geotextile Static Puncture Fixture

UGC-PF-050 Geotextile Static Puncture Fixture

50 kN CBR-type static puncture fixture with sample positioning tool for geotextiles and geogrids, used for ASTM D6241 an

UGC-PF-020 Bi-Metal Coin Core Push-Out Fixture

UGC-PF-020 Bi-Metal Coin Core Push-Out Fixture

Compression fixture that punches the center pill out of a bi-metal coin or token and records the push-out shear force ho

UGC-PF-060 Ball Burst Puncture Fixture ISO 9073-5

UGC-PF-060 Ball Burst Puncture Fixture ISO 9073-5

Ring clamp with 44.45 mm opening and 25.4 mm ball probe for bursting strength of knits, nonwovens and coated fabrics, ra

Dental Ceramic Biaxial Flexure Fixture ISO 6872 FAQ

What is a piston-on-three-balls test?

A disc rests on three balls and is loaded at its center by a small piston, so it bends biaxially. It is the biaxial flexure arrangement used in ISO 6872 for dental ceramics.

 

What are the support and piston dimensions?

Three hardened steel balls of Ø 3.175 mm and a tungsten carbide piston of Ø 1.4 mm.

 

Can it be used inside a temperature chamber?

Yes. The stainless steel fixture is usable from -70 to +280 °C in an environmental chamber.

 

Is there a version for ceramic substrates?

Yes, a version for ASTM F394 modulus-of-rupture testing of ceramic substrates is available. Specify it when ordering.

 

What is the maximum load?

2.5 kN (562 lbf), which suits a Galdabini 2.5 kN benchtop UTM or larger frame.

 

What do I need to specify?

The standard (ISO 6872 or ASTM F394), your disc specimen diameter and the machine coupling.

 

Need help configuring a puncture fixture?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.

bottom of page