Home › Products › Puncture Fixtures › Small Sample Slow Rate Penetration Fixture





Small Sample Slow Rate Penetration Fixture
Manually clamped film penetration fixture based on ASTM F1306, sized for strips up to 50 mm wide or discs up to 64 mm in diameter.
Brand: Universal Grip | Model: UGC-PF-090 | ● Configured to order
Based on ASTM F1306, adapted for small samples
Ø 35 mm (1.375 in) specimen opening with 1.6 mm (0.06 in) probe tip radius
Takes strips up to 50 mm wide or discs up to Ø 64 mm
Manual clamping; no air supply needed
20 kN (4,496 lbf) standard version, 10 kN short-handle version
Nickel-plated steel, 2.1 kg (4.6 lb) including probe
Standards: ASTM F1306
Trusted by Global Companies


























Slow rate penetration fixture for small film samples
The UGC-PF-090 measures the slow rate penetration resistance of flexible barrier films and laminates. It is based on ASTM F1306 but sized for smaller samples, so labs can test material that never exists in full-size sheets: lab-scale extrusion trials, coating development runs, or panels cut from finished pouches and lidding. Packaging R&D groups, film producers and medical packaging teams use it to rank structures early, before committing to larger runs. For the full test method, see our ASTM F1306 guide.
How it works
The film is clamped over a Ø 35 mm (1.375 in) opening with a manual clamp operated by handles. A probe with a 1.6 mm (0.06 in) tip radius, driven by the crosshead at a controlled speed, presses into the film until it penetrates. The machine records force against probe travel up to the point of penetration. The clamp accepts strip samples up to 50 mm (1.97 in) wide or discs up to Ø 64 mm (2.52 in), so you can use whatever sample geometry your material allows.
Versions
Standard: long clamping handles, rated 20 kN (4,496 lbf)
Short handles: compact version, rated 10 kN (2,248 lbf)
The short-handle version is useful where crosshead space or a chamber limits how far the handles can swing. Both connect with a female clevis of Ø 15.9 mm (5/8 in) or Ø 20 mm; other couplings are on request or via adapters. The fixture is nickel-plated steel, weighs 2.1 kg (4.6 lb) including the probe, and covers 0 to +70 °C (32 to 158 °F) as standard.
Manual or pneumatic?
Manual clamping keeps the fixture simple and portable, with no air line to run. If you test full-size F1306 specimens in volume and want clamping force that does not depend on the operator, the UGC-PF-100 pneumatic barrier film penetration fixture clamps with up to 5 kN at 7 bar over the same 35 mm opening.
Typical applications
Barrier films
Food packaging laminates
Medical packaging
Film R&D
Machine fit
Film penetration forces are modest, so a Mark-10 F-series motorized tester (100 to 1,500 lbF) or a Galdabini 2.5 kN or 5 kN benchtop UTM is a typical match.
Tell us your sample format, handle version and machine coupling, and we will configure the fixture: request a quote.
Small Sample Slow Rate Penetration Fixture Specifications
MAX. LOAD
20 kN (standard); 10 kN (short handles)
TEST TYPE
Slow rate penetration
SPECIMEN OPENING
Ø 35 mm (1.375 in)
PROBE
Tip radius 1.6 mm (0.06 in)
SPECIMEN SIZE
Strips max. 50 mm wide; discs max. Ø 64 mm
CLAMPING
Manual, with clamping handles
COUPLING
Female clevis Ø 15.9 mm (5/8 in) or Ø 20 mm; others on request
MATERIAL
Steel, nickel-plated
TEMPERATURE RANGE
0 to +70 °C (others on request)
WEIGHT
2.1 kg incl. probe
DRAWINGS
Dimensional drawings on data sheet; certified drawings and 3D models on request
PART REF.
UGC-PF-090
Models & Options
Standard version — Long clamping handles, 20 kN (Specify at order)
Short handles — Compact version, 10 kN (Specify at order)
Sample format — Strip (max. 50 mm wide) or disc (max. Ø 64 mm) (Specify at order)
Machine coupling — Female clevis Ø 15.9 mm or Ø 20 mm; others (Specify at order)
What's Included
1 manual-clamp penetration fixture with Ø 35 mm opening
Probe with 1.6 mm tip radius
Coupling per your machine, specified at order
Supported Test Standards
ASTM F1306 — Slow rate penetration of flexible barrier films and laminates (method basis; fixture adapted for small samples)
Compatible Grips & Fixtures
UGC-PF-100 Pneumatic Barrier Film Penetration Fixture — Full-size ASTM F1306 with pneumatic clamping
Motorized Tensile Testers — Mark-10 F-series frames for film penetration
2.5 kN Benchtop UTM — Galdabini frame for film and laminate testing
Plastics Testing — Film tensile and tear setups
Adapters — Couplings for other machine connections
“Customer quote”
Related Equipment
UGC-PF-100 Pneumatic Barrier Film Penetration Fixture
Pneumatically clamped fixture for ASTM F1306 slow rate penetration of barrier films, with 5 kN clamping at 7 bar for sli
UGC-PF-080 Packaging Puncture Resistance Fixture EN 14477
Needle puncture fixture with interchangeable 0.8 mm round or 3 mm flat needles for packaging films, pouches and laminate
UGC-PF-070 Leather Ball Burst Fixture ISO 3379
Compact 2 kN ball burst fixture that records distension at grain crack and at burst, showing whether leather will surviv
UGC-PF-110 Tissue Paper Ball Burst Fixture ISO 12625-9
Stainless steel ball burst fixture with water tray for dry and wet tissue, pneumatic (500 N) or manual (10 kN) clamping,
Small Sample Slow Rate Penetration Fixture FAQ
Is this fixture for ASTM F1306?
It is based on ASTM F1306 and keeps the 35 mm opening and 1.6 mm probe radius, but is adapted for smaller samples. Our ASTM F1306 guide covers the method.
What sample sizes fit?
Strips up to 50 mm (1.97 in) wide or discs up to Ø 64 mm (2.52 in).
What is the difference between the two versions?
The standard version has long clamping handles and is rated 20 kN; the compact short-handle version is rated 10 kN.
Does it need compressed air?
No. Clamping is manual. For pneumatic clamping, see the UGC-PF-100.
Which couplings are available?
Female clevis Ø 15.9 mm (5/8 in) or Ø 20 mm; others on request or via adapters.
Which machine suits it?
Film penetration forces are modest, so a Mark-10 F-series motorized tester or a Galdabini 2.5 or 5 kN benchtop UTM is typical.