Pneumatic Barrier Film Penetration Fixture – ASTM F1306
Pneumatically clamped fixture for ASTM F1306 slow rate penetration of barrier films, with 5 kN clamping at 7 bar for slip-free, repeatable tests.
Brand: Universal Grip | Model: UGC-PF-100 | ● Configured to order
Pneumatic clamping: 5 kN (1,124 lbf) at 7 bar, 9 kN at 12 bar
Ø 35 mm (1.375 in) specimen opening, 1.6 mm (0.06 in) probe tip radius
Clamping force does not depend on the operator
Max. vertical test load 3 kN (674 lbf)
EN 14477 and large-opening (Ø 45 mm) versions available
Nickel-plated steel, 5.2 kg (11.5 lb)
Standards: ASTM F1306 · EN 14477
Trusted by Global Companies


























Pneumatic penetration fixture for flexible barrier films
The UGC-PF-100 measures the slow rate penetration resistance of flexible barrier films and laminates according to ASTM F1306. Food and medical packaging labs, film producers and pouch converters use the test to show how well a structure resists being pierced by a slowly advancing point, such as a sharp edge of the product pressing outward. In this test, film slipping in the clamp is the most common source of scatter, and the PF-100 tackles it with pneumatic clamping. See our complete ASTM F1306 guide for specimen preparation, procedure and calculations.
How it works
A pneumatic cylinder clamps the film over a Ø 35 mm (1.375 in) opening with 5 kN (1,124 lbf) at 7 bar (102 psi), or up to 9 kN (2,023 lbf) at 12 bar (174 psi). A probe with a 1.6 mm (0.06 in) tip radius then penetrates the film at a controlled speed, and the machine records force and displacement to penetration. The clamping force is set by air pressure, not by how hard an operator turns a handle, so it stays the same from specimen to specimen and from shift to shift. The maximum vertical (test) load on the fixture is 3 kN (674 lbf), which is separate from the clamping force.
Versions
ASTM F1306: Ø 35 mm opening, 1.6 mm probe radius
EN 14477: Ø 10 mm opening, clamping 5 kN at 7 bar (11 kN, or 2,473 lbf, at 16 bar / 232 psi)
Large opening: Ø 45 mm ring, 2 mm probe radius
When you order, tell us the standard and opening size, the air pressure available in your lab and your machine coupling. The standard connection is a female clevis with a Ø 15.9 mm (5/8 in) pin; others are on request or via adapters. The fixture is nickel-plated steel, weighs 5.2 kg (11.5 lb) and covers 0 to +70 °C (32 to 158 °F) as standard.
Typical applications
Barrier films and laminates
Food and medical packaging
Flexible pouches
Film quality control
Machine fit
With a 3 kN vertical rating, the fixture suits a Mark-10 F-series motorized tester (up to 1,500 lbF) or a Galdabini 5 kN benchtop UTM. For small samples without an air supply, the UGC-PF-090 uses the same opening and probe radius with a manual clamp; for needle puncture to EN 14477, see the UGC-PF-080. Other pneumatic setups are on our pneumatic grips page.
Tell us your standard, opening size, air pressure and machine coupling, and we will configure the fixture: request a quote.
Pneumatic Barrier Film Penetration Fixture Specifications
MAX. VERTICAL LOAD
3 kN
CLAMPING FORCE
5 kN at 7 bar (9 kN at 12 bar)
TEST TYPE
Slow rate penetration
SPECIMEN OPENING
Ø 35 mm (1.375 in)
PROBE
Tip radius 1.6 mm (0.06 in)
OPERATION
Pneumatic clamping
COUPLING
Female clevis, Ø 15.9 mm (5/8 in) pin; others on request
MATERIAL
Steel, nickel-plated
TEMPERATURE RANGE
0 to +70 °C (others on request)
WEIGHT
5.2 kg
DRAWINGS
Dimensional drawings on data sheet; certified drawings and 3D models on request
PART REF.
UGC-PF-100
Models & Options
ASTM F1306 version — Ø 35 mm opening, 1.6 mm probe radius (Specify at order)
EN 14477 version — Ø 10 mm opening, clamping 5 kN at 7 bar (11 kN at 16 bar) (Specify at order)
Large opening version — Ø 45 mm ring, 2 mm probe radius (Specify at order)
Machine coupling — Female clevis Ø 15.9 mm standard; others (On request)
What's Included
1 pneumatic penetration fixture with clamping cylinder
Specimen opening and probe for the version ordered
Compressed air supply: specify the pressure available in your lab at order
Supported Test Standards
ASTM F1306 — Slow rate penetration resistance of flexible barrier films and laminates
EN 14477 — Packaging, flexible packaging material, puncture resistance (version)
Compatible Grips & Fixtures
UGC-PF-090 Small Sample Slow Rate Penetration Fixture — Manual clamp, small samples
UGC-PF-080 Packaging Puncture Resistance Fixture — Needle puncture to EN 14477
Motorized Tensile Testers — Mark-10 F-series frames for the 3 kN rating
5 kN Benchtop UTM — Galdabini frame for film testing
Pneumatic Grips — Air-operated grips for film tensile tests
“Customer quote”
Related Equipment
UGC-PF-080 Packaging Puncture Resistance Fixture EN 14477
Needle puncture fixture with interchangeable 0.8 mm round or 3 mm flat needles for packaging films, pouches and laminate
UGC-PF-090 Small Sample Slow Rate Penetration Fixture
Manually clamped film penetration fixture based on ASTM F1306, sized for strips up to 50 mm wide or discs up to 64 mm in
UGC-PF-110 Tissue Paper Ball Burst Fixture ISO 12625-9
Stainless steel ball burst fixture with water tray for dry and wet tissue, pneumatic (500 N) or manual (10 kN) clamping,
UGC-PF-120 Fabric Stiffness Puncture Fixture
Circular bend fixture with 38.1 mm orifice and 25.4 mm flat plunger that measures fabric stiffness in all directions at
Pneumatic Barrier Film Penetration Fixture FAQ
Which standard is the UGC-PF-100 for?
ASTM F1306 slow rate penetration of flexible barrier films and laminates. Our ASTM F1306 guide covers specimens, procedure and calculations.
Why pneumatic clamping?
Air pressure sets the clamping force, so it is the same for every specimen and every operator, which cuts scatter from film slipping in the clamp.
How much clamping force does it apply?
5 kN at 7 bar (1,124 lbf at 102 psi), or up to 9 kN at 12 bar (174 psi).
What is the maximum test load?
3 kN (674 lbf) vertical load, separate from the clamping force.
Are other openings available?
Yes: an EN 14477 version with a Ø 10 mm opening and a large-opening version with a Ø 45 mm ring and 2 mm probe radius.
What do I need to tell you to order?
The standard and opening size, the air pressure available in your lab and your machine coupling.