top of page

Home  ›  Products  ›  Peel Fixtures  ›  Climbing Drum Peel Fixture

UGC-PL-090 climbing drum peel fixture with flanged aluminum drum and loading straps by Universal Grip
UGC-PL-090 climbing drum peel fixture with flanged aluminum drum and loading straps by Universal Grip
UGC-PL-090 Climbing Drum Peel Fixture by Universal Grip, view 2
UGC-PL-090 Climbing Drum Peel Fixture by Universal Grip, view 3
UGC-PL-090 Climbing Drum Peel Fixture by Universal Grip, view 4

Climbing Drum Peel Fixture – ASTM D1781 and EN 2243-3

10 kN climbing drum peel fixture for sandwich panels and bonded laminates, with 102 mm (ASTM D1781) and 100 mm (EN 2243-3) drums.

Brand: Universal Grip  |  Model: UGC-PL-090  |  ● Configured to order

  • Flanged drum with flexible loading straps rolls up the specimen and peels the facing

  • Rated to 10 kN (about 2,248 lbf)

  • 102 mm (4 in) drum for ASTM D1781; 100 mm drum for EN 2243-3 and DIN 53295

  • Lightweight aluminum drum with steel straps and fittings; 3.4 kg drum

  • Optional adjustable table helps start the test

  • Quick-release drum clamp version available

Standards: ASTM D1781  ·  EN 2243-3  ·  DIN 53295

Trusted by Global Companies

Climbing drum peel fixture for sandwich panels

The UGC-PL-090 measures the peel resistance of adhesive bonds between a flexible facing and a rigid core or adherend. It is the fixture for the climbing drum peel test, used to qualify honeycomb sandwich panels, metal-faced laminates and aerospace bonded structures. The fixture is rated to 10 kN (about 2,248 lbf) and is offered with drum diameters for ASTM D1781 and for EN 2243-3 and DIN 53295.

 

How it works

The free end of the flexible facing is clamped to a flanged drum. Flexible loading straps wrap around the flanges and connect to the testing machine. As the machine pulls the straps, the drum rolls up the specimen, winding the facing onto itself and peeling it away from the core. Because the drum and flange geometry are fixed, results from specimen to specimen and lab to lab can be compared directly.

See our complete ASTM D1781 guide for specimen preparation, procedure and calculations; this page covers the hardware.

 

Versions and accessories

  • ASTM D1781 version: 102 mm (4 in) drum, for laminate specimens 25 mm (about 1 in) wide and sandwich specimens 76 mm (3 in) wide, at least 254–305 mm (10–12 in) long.

  • EN 2243-3 / DIN 53295 version: 100 mm drum for specimens 75 x 300 mm.

  • Adjustable table: an accessory that helps start the test.

  • Quick-release version: drum clamp with a quick-release lever for faster specimen changes.

The drum is aluminum, with steel straps and fittings. The drum weighs 3.4 kg (about 7.5 lb), or 4.35 kg (about 9.6 lb) in total with adapter. The coupling is a female clevis for a Ø 15.9 mm (5/8 in) or Ø 20 mm pin, with others on request, and the operating range is 0 to +70 °C, with other ranges on request.

 

Machine requirements

The climbing drum travels up the full length of the specimen, so the test needs at least 1200 mm (about 47 in) of machine travel. Check this before you order; it often points to a floor-standing universal testing machine rather than a compact benchtop frame. The 10 kN rating matches Galdabini universal testing machines in that class, and other connections are handled with our adapters.

 

Upper grip

The fixture is supplied as one unit and the upper grip for the specimen is ordered separately. Mechanical wedge grips are a common choice for metal-faced panels.

Tell us the standard you test to (drum diameter), your machine coupling, the upper grip and your available machine travel, and we will configure the fixture. Request a quote.

Product video

At a glance

  • Max. load: 10 kN (approx. 2,248 lbf)

  • Test type: Climbing drum peel

  • Drum diameter: 102 mm (4 in) or 100 mm, by version

  • Specimen size: 25 or 76 mm wide x min. 254–305 mm (ASTM); 75 x 300 mm (EN)

  • Min. machine travel: 1200 mm (approx. 47 in)

  • Coupling: Female clevis, Ø 15.9 mm (5/8 in) or Ø 20 mm pin; others on request

Climbing Drum Peel Fixture Specifications

MAX. LOAD

10 kN (approx. 2,248 lbf)

 

TEST TYPE

Climbing drum peel

 

DRUM DIAMETER

102 mm (4 in) or 100 mm, by version

 

SPECIMEN SIZE

25 or 76 mm wide x min. 254–305 mm (ASTM); 75 x 300 mm (EN)

 

MIN. MACHINE TRAVEL

1200 mm (approx. 47 in)

 

COUPLING

Female clevis, Ø 15.9 mm (5/8 in) or Ø 20 mm pin; others on request

 

MATERIAL

Aluminum drum, steel components

 

TEMPERATURE RANGE

0 to +70 °C; other ranges on request

 

WEIGHT

3.4 kg drum; 4.35 kg total with adapter

 

SCOPE OF SUPPLY

1 fixture; upper grip ordered separately

 

STANDARDS NAMED

ASTM D1781, EN 2243-3, DIN 53295

 

DRAWINGS

Dimensional drawings on page 2 of the data sheet; certified drawings and 3D models on request

 

PART REF.

UGC-PL-090

 

Models & Options

  • ASTM D1781 version — 102 mm drum; laminates 25 mm wide, sandwich 76 mm wide (Specify at order)

  • EN 2243-3 / DIN 53295 version — 100 mm drum; specimen 75 x 300 mm (Specify at order)

  • Adjustable table — Accessory that helps start the test (Specify at order)

  • Quick-release version — Drum clamp with quick-release lever (Specify at order)

  • Upper grip — Grip for the specimen (On request)

What's Included

  • 1 climbing drum peel fixture: flanged aluminum drum with steel loading straps and fittings

  • Drum diameter for the standard ordered (102 mm or 100 mm)

  • Female clevis coupling for Ø 15.9 mm or Ø 20 mm pin

  • Upper grip ordered separately; adjustable table optional

Supported Test Standards

  • ASTM D1781 — Climbing drum peel for adhesives

  • EN 2243-3 — Aerospace structural adhesives – climbing drum peel of metal-to-honeycomb core

  • DIN 53295 — Climbing drum peel test of sandwich constructions

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-PL-080 Floating Roller Peel Fixture

UGC-PL-080 Floating Roller Peel Fixture

10 kN floating roller peel fixture for flexible-to-rigid adhesive bonds, with ASTM D3167 and EN/ISO roller geometries.

UGC-PL-100 90 Degree Peel Grip 25 mm Roller

UGC-PL-100 90 Degree Peel Grip 25 mm Roller

Compact 500 N peel grip that holds the rigid substrate under a height-adjustable Ø 25 mm roller for 90° peel tests on st

UGC-PL-070 ID Card 90 Degree Peel Fixture ISO IEC 10373

UGC-PL-070 ID Card 90 Degree Peel Fixture ISO IEC 10373

Roller-guided 90° peel fixture, rated to 2 kN, for testing the adhesion of layers on identification cards to ISO/IEC 103

UGC-PL-110 Floor and Wall Covering Peel Fixture EN 1372

UGC-PL-110 Floor and Wall Covering Peel Fixture EN 1372

Floating roller peel fixture for floor and wall covering adhesives to EN 1372 and ISO 22631, with 57, 97 or 157 mm rolle

Climbing Drum Peel Fixture FAQ

What does the climbing drum peel test measure?

The peel resistance of the bond between a flexible facing and a rigid core or adherend. Our ASTM D1781 guide covers the procedure and calculations.

 

Which drum diameter do I need?

102 mm (4 in) for ASTM D1781, or 100 mm for EN 2243-3 and DIN 53295.

 

How much machine travel is required?

At least 1200 mm (about 47 in), because the drum climbs the full specimen length. Confirm your machine travel before ordering.

 

What specimen sizes fit?

For ASTM, laminates 25 mm wide and sandwich specimens 76 mm wide, at least 254–305 mm long; for EN, specimens 75 x 300 mm.

 

Is the upper grip included?

No. The fixture is supplied as one unit and the upper grip is ordered separately; an adjustable table to help start the test is optional.

 

What load and temperature range apply?

10 kN (about 2,248 lbf) and 0 to +70 °C as standard; other ranges on request. Contact us to configure a setup.

 

Need help configuring a peel fixture?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.

bottom of page