Carrier fixture for 180° peel adhesion tests on self-adhesive tapes to EN 1939, ISO 29862 and FINAT methods, with steel, stainless or glass plates.

Brand: Universal Grip  |  Model: UGC-PL-140  |  ● Configured to order

  • 180° peel geometry for self-adhesive tapes, labels and protective films

  • Interchangeable test plates: stainless steel 1.4301, plain steel or float glass

  • Takes specimens up to 50 mm wide and 300 mm long (1.97 x 11.8 in)

  • Carrier rated for 10 kN (2,248 lbf)

  • Nickel-plated steel carrier, 2.4 kg

  • Female clevis coupling with Ø 15.9 mm (5/8 in) or Ø 20 mm pin

Standards: ISO 29862  ·  EN 1939  ·  FINAT FTM 1  ·  FINAT FTM 3

Trusted by Global Companies

180° peel carrier for self-adhesive tapes and labels

The UGC-PL-140 is a carrier fixture for 180° peel adhesion tests on self-adhesive tapes, labels, protective films and other pressure-sensitive adhesives. It is built for tests to EN 1939, ISO 29862 and FINAT FTM 1, 3 and 4, giving tape manufacturers and converting QC labs one fixture that moves between methods by changing the test plate. See our complete ISO 29862 guide for specimen preparation, procedure and calculations.

 

How it works

The tape is applied to a test plate that sits vertically in the nickel-plated steel carrier. Its free end is folded back over itself and clamped in an upper grip, so as the crosshead rises the tape peels away from the plate at 180°. The carrier takes specimens up to 50 mm (1.97 in) wide and 300 mm (11.8 in) long.

 

Test plates for each method

  • Stainless steel 1.4301 – Ra 0.05 µm ±0.025, for EN 1939 and ISO 29862

  • Plain steel – untreated, for FINAT FTM 3

  • Float glass – for FINAT FTM 1, 3 and 4

Plates are interchangeable and ordered separately, so you can stock the surfaces your customers specify and change between them in the same carrier.

 

Ratings and coupling

The carrier is rated for 10 kN (2,248 lbf), far above typical tape peel forces, and weighs 2.4 kg. It connects through a female clevis with a Ø 15.9 mm (5/8 in) or Ø 20 mm pin, with other couplings on request. The standard working range is 0 to +70 °C, and other ranges are available on request.

 

Typical applications

  • Self-adhesive tapes

  • Labels and label stock

  • Protective films

  • Pressure-sensitive adhesives

 

Specimen application and machine fit

Consistent rolling pressure matters as much as the fixture, so application rollers are available: our 2 kg peel test sample roller (UGC-PL-200) is offered in standard and FINAT versions. Tape peel forces fit a Mark-10 F-series motorized tester as well as our Galdabini benchtop universal testing machines, and adapters cover other frames. The upper grip for the free tape end is ordered separately.

Let us know which standard you test to, the plate surfaces you need and your machine coupling, and we'll put the carrier, plates, roller and upper grip into one quote.

Product video

At a glance

  • Max. load: 10 kN (2,248 lbf)

  • Test type: 180° peel

  • Specimen size: Max. 50 mm wide x 300 mm long

  • Test plates: Stainless steel 1.4301, plain steel or float glass

  • Stainless plate finish: Ra 0.05 µm ±0.025

  • Coupling: Female clevis Ø 15.9 mm (5/8 in) or Ø 20 mm pin; others on request

180 Degree Tape Peel Fixture EN 1939 Specifications

MAX. LOAD

10 kN (2,248 lbf)

 

TEST TYPE

180° peel

 

SPECIMEN SIZE

Max. 50 mm wide x 300 mm long

 

TEST PLATES

Stainless steel 1.4301, plain steel or float glass

 

STAINLESS PLATE FINISH

Ra 0.05 µm ±0.025

 

COUPLING

Female clevis Ø 15.9 mm (5/8 in) or Ø 20 mm pin; others on request

 

BODY MATERIAL

Steel, nickel-plated

 

TEMPERATURE RANGE

0 to +70 °C; other ranges on request

 

WEIGHT

2.4 kg (carrier)

 

SCOPE OF SUPPLY

1 carrier; test plates and upper grip ordered separately

 

PART REF.

UGC-PL-140

 

Models & Options

  • Carrier — 180° peel carrier, supplied without test plate (Specify at order)

  • Stainless steel plate — 1.4301, Ra 0.05 µm ±0.025, for EN 1939 and ISO 29862 (Specify at order)

  • Plain steel plate — Untreated, for FINAT FTM 3 (Specify at order)

  • Float glass plate — For FINAT FTM 1, 3 and 4 (Specify at order)

  • Application roller — 2 kg hand roller for applying specimens (UGC-PL-200) (Specify at order)

  • Other couplings — Alternatives to the clevis couplings to suit your machine (On request)

What's Included

  • 1 nickel-plated steel 180° peel carrier

  • Test plates ordered separately in the material your method calls for

  • Upper grip for the free tape end ordered separately

  • Application rollers available as an add-on

Supported Test Standards

  • ISO 29862 — Self-adhesive tapes – determination of peel adhesion properties

  • EN 1939 — European method for peel adhesion of self-adhesive tapes

  • FINAT FTM 1 — 180° peel adhesion of self-adhesive materials

  • FINAT FTM 3 — Low-speed peel adhesion of self-adhesive materials

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-PL-200 Peel Test Sample Roller

UGC-PL-200 Peel Test Sample Roller

2 kg polyurethane-coated hand roller that applies tape specimens to peel test panels with even pressure, in ASTM, EN/ISO

UGC-PL-130 German Wheel Peel Fixture

UGC-PL-130 German Wheel Peel Fixture

Free-rotating wheel fixture that holds a constant 90° peel angle on tapes, films, release coatings and flexible circuits

UGC-PL-150 90 Degree Adhesive Peel Fixture ISO 8510-1

UGC-PL-150 90 Degree Adhesive Peel Fixture ISO 8510-1

Wire-driven sliding-table fixture that keeps the peel point centered at a constant 90° for flexible-to-rigid adhesive bo

UGC-PL-120 90 Degree Adhesive Peel Grip DIN 40802

UGC-PL-120 90 Degree Adhesive Peel Grip DIN 40802

Six-roller 90° peel fixture that measures the adhesion of flexible films to rigid substrates up to 7 mm thick, rated 2 k

180 Degree Tape Peel Fixture EN 1939 FAQ

Which test plate do I need?

It depends on the method: stainless steel 1.4301 for EN 1939 and ISO 29862, untreated plain steel for FINAT FTM 3, and float glass for FINAT FTM 1, 3 and 4.

 

Are test plates and the upper grip included?

No. The fixture is supplied as one carrier; test plates and the upper grip are ordered separately so you only buy the surfaces you use.

 

What specimen size fits?

Specimens up to 50 mm (1.97 in) wide and 300 mm (11.8 in) long.

 

Where can I find the ISO 29862 procedure?

See our ISO 29862 guide for specimen preparation, test procedure and calculations. This page covers the fixture hardware.

 

Do I need an application roller?

Tape peel methods specify a defined rolling pressure when applying specimens. Our 2 kg peel test sample roller is available in standard and FINAT versions.

 

Why is the carrier rated for 10 kN?

The 10 kN (2,248 lbf) rating gives a large safety margin over typical tape peel forces, so the carrier can also run on larger frames without concern.

 

Need help configuring a peel fixture?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.