top of page

Home  ›  Products  ›  Peel Fixtures  ›  90 Degree Adhesive Peel Fixture ISO 8510-1

UGC-PL-150 sliding table 90 degree adhesive peel fixture for ISO 8510-1 tests by Universal Grip
UGC-PL-150 sliding table 90 degree adhesive peel fixture for ISO 8510-1 tests by Universal Grip
UGC-PL-150 90 Degree Adhesive Peel Fixture ISO 8510-1 by Universal Grip, view 2
UGC-PL-150 90 Degree Adhesive Peel Fixture ISO 8510-1 by Universal Grip, view 3
UGC-PL-150 90 Degree Adhesive Peel Fixture ISO 8510-1 by Universal Grip, view 4

90 Degree Adhesive Peel Fixture ISO 8510-1 – Sliding Table

Wire-driven sliding-table fixture that keeps the peel point centered at a constant 90° for flexible-to-rigid adhesive bonds, rated 5 kN.

Brand: Universal Grip  |  Model: UGC-PL-150  |  ● Configured to order

  • Wire-linked sliding table keeps a constant 90° peel angle through the test

  • Peel point stays centered under the load axis

  • Fixed 90° or adjustable-angle versions in standard and 130 mm widths

  • Specimens up to 50 x 280 mm, or 100 x 280 mm on the wide fixture

  • Rated for 5 kN (1,124 lbf); nickel-plated steel body

  • Supplied with mounting bar, wire and wire end tubes

Standards: ISO 8510-1  ·  ASTM D6862  ·  ISO 29862  ·  ASTM D3330

Trusted by Global Companies

Sliding-table 90° peel fixture for flexible-to-rigid adhesive bonds

The UGC-PL-150 measures the peel strength of a flexible adherend bonded to a rigid substrate at a controlled 90° angle. It is used for tests such as ISO 8510-1 and ASTM D6862, and the same hardware is listed for tape peel methods including ISO 29862 and ASTM D3330. Adhesive formulators, tape producers and labs testing coatings on metal or thermoplastic elastomer bonds choose it when a constant angle and a centered peel point matter. See our complete ISO 8510-1 guide for specimen preparation, procedure and calculations.

 

How it works

The rigid substrate is mounted on a sliding table. A wire links the table to the crosshead, so as the crosshead lifts the flexible layer the table travels with it. The peel point therefore stays centered under the load axis and the angle stays constant from start to finish, instead of drifting as the bond line is used up. Each fixture comes with its mounting bar, wire and wire end tubes.

 

Versions

  • 90°, standard width – specimens up to 50 x 280 mm, 5.4 kg

  • Adjustable angle, standard width – adjustable peel angle 60°, 7.6 kg

  • 90°, 130 mm wide – specimens up to 100 x 280 mm, 7.1 kg

  • Adjustable angle, 130 mm wide – ±30° peel angle, 9.2 kg

The 130 mm figure is the fixture width; it takes specimens up to 100 mm (3.9 in) wide. Stainless steel carrier plates, 2 mm thick, come in sets of 10: 56 x 280 mm plates and 106 x 280 mm plates for the 130 mm fixtures.

 

Ratings and coupling

All versions are rated for 5 kN (1,124 lbf) and built from nickel-plated steel. The fixture connects through an M6 thread at the top and a female clevis with a Ø 15.9 mm (5/8 in) pin at the bottom; other couplings are available on request. The standard working range is 0 to +70 °C, with other ranges on request.

 

Typical applications

  • Structural and flexible adhesives

  • Adhesive tapes

  • Coatings on metal substrates

  • Thermoplastic elastomer bonds

 

Choosing the right setup

An upper grip for the flexible adherend is required and ordered separately. Our Galdabini 10 kN universal testing machine covers the full 5 kN rating, and a Mark-10 F-series motorized tester suits lower-force tape and film peels. For tape specimens, add our peel test sample roller (UGC-PL-200).

Tell us your specimen width, whether you need a fixed or adjustable angle, and your machine coupling, and we'll configure the fixture, carrier plates and upper grip for your quote.

Product video

At a glance

  • Max. load: 5 kN (1,124 lbf)

  • Test type: 90° peel, sliding table

  • Specimen size: Max. 50 x 280 mm (standard) or 100 x 280 mm (wide)

  • Peel angle: 90°; adjustable versions 60° or ±30°

  • Coupling: M6 thread (upper) and female clevis Ø 15.9 mm (5/8 in) pin (lower); others on request

  • Body material: Steel, nickel-plated

90 Degree Adhesive Peel Fixture ISO 8510-1 Specifications

MAX. LOAD

5 kN (1,124 lbf)

 

TEST TYPE

90° peel, sliding table

 

SPECIMEN SIZE

Max. 50 x 280 mm (standard) or 100 x 280 mm (wide)

 

PEEL ANGLE

90°; adjustable versions 60° or ±30°

 

COUPLING

M6 thread (upper) and female clevis Ø 15.9 mm (5/8 in) pin (lower); others on request

 

BODY MATERIAL

Steel, nickel-plated

 

CARRIER PLATES

Stainless steel, 2 mm thick, 56 x 280 mm or 106 x 280 mm, sets of 10

 

TEMPERATURE RANGE

0 to +70 °C; other ranges on request

 

WEIGHT

5.4 kg to 9.2 kg, by version

 

SCOPE OF SUPPLY

Fixture with mounting bar, wire and wire end tubes; upper grip ordered separately

 

PART REF.

UGC-PL-150

 

Models & Options

  • 90°, standard width — Specimens up to 50 mm wide, 5.4 kg (Specify at order)

  • Adjustable angle, standard width — Adjustable peel angle 60°, 7.6 kg (Specify at order)

  • 90°, 130 mm wide — Specimens up to 100 mm wide, 7.1 kg (Specify at order)

  • Adjustable angle, 130 mm wide — ±30° peel angle, 9.2 kg (Specify at order)

  • Carrier plates 56 x 280 mm — Stainless steel, 2 mm thick, set of 10 (Specify at order)

  • Carrier plates 106 x 280 mm — Stainless steel, 2 mm thick, for 130 mm fixtures, set of 10 (Specify at order)

  • Other couplings — Alternatives to the M6 / Ø 15.9 mm clevis connections (On request)

What's Included

  • Sliding-table 90° peel fixture in the width and angle version ordered

  • Mounting bar, wire and wire end tubes

  • Carrier plates ordered separately in sets of 10

  • Upper grip for the flexible adherend ordered separately

Supported Test Standards

  • ISO 8510-1 — Adhesives – 90° peel test for flexible-to-rigid bonded assemblies

  • ASTM D6862 — 90 degree peel resistance of adhesives

  • ISO 29862 — Self-adhesive tapes – determination of peel adhesion properties

  • ASTM D3330 — Peel adhesion of pressure-sensitive tape

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-PL-200 Peel Test Sample Roller

UGC-PL-200 Peel Test Sample Roller

2 kg polyurethane-coated hand roller that applies tape specimens to peel test panels with even pressure, in ASTM, EN/ISO

UGC-PL-140 180 Degree Tape Peel Fixture EN 1939

UGC-PL-140 180 Degree Tape Peel Fixture EN 1939

Carrier fixture for 180° peel adhesion tests on self-adhesive tapes to EN 1939, ISO 29862 and FINAT methods, with steel,

UGC-PL-160 90 Degree Rubber Adhesion Peel Fixture ISO 813

UGC-PL-160 90 Degree Rubber Adhesion Peel Fixture ISO 813

Articulated 90° peel fixture for rubber-to-rigid substrate adhesion tests such as ISO 813 and ASTM D429, rated 5 kN, in

UGC-PL-130 German Wheel Peel Fixture

UGC-PL-130 German Wheel Peel Fixture

Free-rotating wheel fixture that holds a constant 90° peel angle on tapes, films, release coatings and flexible circuits

90 Degree Adhesive Peel Fixture ISO 8510-1 FAQ

How does the sliding table keep the angle at 90°?

A wire links the table to the crosshead, so the table moves by the same distance the flexible layer is lifted. The peel point stays centered under the load axis for the whole test.

 

Where can I find the ISO 8510-1 test procedure?

See our ISO 8510-1 guide for specimen preparation, procedure and calculations. This page covers the fixture hardware and its versions.

 

Should I choose the standard or the 130 mm wide fixture?

The standard fixture takes specimens up to 50 x 280 mm. The 130 mm wide fixture takes specimens up to 100 x 280 mm and uses the 106 x 280 mm carrier plates.

 

What do the adjustable-angle versions offer?

They let you run peels away from 90°: the standard-width version has an adjustable peel angle of 60° and the 130 mm version ±30°. Tell us the angles you need and we'll confirm the right version.

 

Is an upper grip included?

No. The fixture comes with its mounting bar, wire and wire end tubes; the upper grip for the flexible adherend is ordered separately, for example from our tensile grips.

 

Which testing machine is a good match?

Our 10 kN universal testing machine covers the full 5 kN (1,124 lbf) rating. The upper coupling is an M6 thread and the lower a female clevis with Ø 15.9 mm pin.

 

Need help configuring a peel fixture?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.

bottom of page