Backlash-free 30 kN fatigue grips, about 1.5 kg each, for force-controlled tests on acrylic bone cement and other flat or round rigid specimens.
Brand: Universal Grip | Model: UGC-VG-160 | ● Configured to order
Rated 30 kN (6,744 lbf) for static and dynamic loading
Backlash-free clamping keeps the specimen tight through load reversals
About 1.5 kg (3.3 lb) per grip keeps moving mass low
Serrated 30 x 30 mm jaws for flat specimens 0–12 mm (0 to 0.47 in) thick
V-jaw version for round specimens Ø 4–12 mm
Hardened 58 HRC nickel-coated steel for -40 to +280 °C
M12 x 1.75 threaded stud coupling for a rigid, aligned load train
Standards: ASTM F2118
Trusted by Global Companies


























Fatigue grips for acrylic bone cement and rigid specimens
The UGC-VG-160 is a compact, lightweight pair of fatigue grips built for constant-amplitude, force-controlled tests on acrylic bone cement to ASTM F2118. Bone cement specimens are small and brittle, and a fatigue test can run for a very large number of cycles in tension and compression. Any looseness in the grips shows up as a dead band each time the load crosses zero, and any excess mass on the actuator limits the frequency you can run. This fixture addresses both: it clamps without backlash and weighs about 1.5 kg (3.3 lb) per grip. Orthopedic device makers, cement suppliers and biomechanics labs are the core users.
How it works
The grips are adjusted with a backlash-free mechanism, so once the specimen is clamped there is no free play to take up when the load reverses from tension to compression. High clamping force holds the specimen without relying on the test load to tighten the jaws. Grips and jaws are made of steel hardened to 58 HRC and nickel-coated, rated from -40 to +280 °C, and each grip mounts on an M12 x 1.75 threaded stud for a stiff, aligned load train.
Flat or round specimens
Flat-specimen version: serrated 30 x 30 mm jaws (1.2 x 45° serration), opening 0–12 mm, set up for ASTM F2118 specimens
Round-specimen version: V-jaws for Ø 4–12 mm with a 1 mm tooth pitch
Diamond-coated jaws: 30 x 30 mm, opening 0–10 mm, for hard or polished surfaces
Beyond bone cement
The same grips suit fatigue work on medical implant materials, steel and aluminum, and CFRP and fiberglass composites, including in temperature chambers within the -40 to +280 °C range. Special builds extend the concept: high-capacity fatigue grips at 50, 100 or 200 kN, and a 30 kN high-temperature version for up to +600 °C with a TiN coating, both on request.
Machine fit
Fatigue tests to ASTM F2118 need a dynamic test system, so these grips are usually fitted to the customer's servo-hydraulic or electrodynamic frame. If its load cell or actuator uses a thread other than M12 x 1.75, we supply adapters. For chamber work see our environmental chambers, and for heavier composite fatigue loads compare the UGC-VG-060 dynamic vise grip rated 100 kN. Other cyclic-loading fixtures are grouped under fatigue grips.
Tell us whether your specimens are flat or round, their size, the capacity and test temperature, and your coupling thread, and we will configure the pair: request a quote.
Bone Cement Fatigue Test Fixture Specifications
MAX. LOAD
30 kN (static and dynamic)
TEST TYPE
Tension-compression fatigue
SPECIMEN SIZE
Flat 0–12 mm thick or round Ø 4–12 mm
CLAMPING SURFACE
30 x 30 mm, serration 1.2 x 45°
COUPLING
M12 x 1.75 threaded stud
MATERIAL
Steel, hardened 58 HRC, nickel-coated
TEMPERATURE RANGE
-40 to +280 °C
WEIGHT
1.5–1.54 kg per grip
SCOPE OF SUPPLY
1 pair of grips
STANDARD
ASTM F2118
PART REF.
UGC-VG-160
Models & Options
Flat-specimen version — Serrated jaws, opening 0–12 mm, for ASTM F2118 specimens (Specify at order)
Round-specimen version — V-jaws for Ø 4–12 mm, 1 mm tooth pitch (Specify at order)
Diamond-coated jaws — 30 x 30 mm, opening 0–10 mm (Specify at order)
High-capacity fatigue grips — 50, 100 or 200 kN special versions (On request)
High-temperature special — 30 kN, up to +600 °C, TiN-coated (On request)
Thread adapter — For load cells and actuators other than M12 x 1.75 (On request)
What's Included
1 pair of fatigue grips with M12 x 1.75 threaded studs
Serrated jaws (flat version) or V-jaws (round version) as ordered
Diamond-coated jaws and thread adapters on request
Supported Test Standards
ASTM F2118 — Constant-amplitude, force-controlled fatigue of acrylic bone cement
Compatible Grips & Fixtures
Fatigue Grips — Other grips for cyclic and dynamic testing
Dynamic Vise Grip for Composites 100 kN — Bolt-clamped fatigue grips for higher loads
Environmental Chambers — Chamber testing within -40 to +280 °C
Adapters — Thread adapters for actuators and load cells
Compression Fixtures — For static compression tests on bone cement
“Customer quote”
Related Equipment
UGC-VG-060 Dynamic Vise Grip for Composites 100 kN
Bolt-clamped 100 kN vise grips for static and tension-tension fatigue tests on polymer composites, with chamber and wate
UGC-VG-150 Fine Wire and Narrow Band Grip 2.5 kN
2.5 kN screw action grip with quick-change 37 x 15 mm insert jaws closing 0–5 mm, for fine wire, thin rope, narrow bands
UGC-VG-170 Textile and Geotextile Vise Grip 50 kN
Vise grip with three screws per side and 40 x 200 mm wave or pyramid jaws, 100 kN clamping at 120 Nm, for textiles, geot
UGC-VG-140 Screw Action Grip 1 kN
Light 1 kN screw action grip with self-tightening articulating jaws and 5 kN clamping at 4 Nm, for films, foils, paper,
Bone Cement Fatigue Test Fixture FAQ
Which test is the UGC-VG-160 built for?
Constant-amplitude, force-controlled fatigue tests on acrylic bone cement to ASTM F2118.
Why does backlash matter in a fatigue grip?
In tension-compression cycling, any play in the grip is taken up every time the load crosses zero, which distorts the waveform and can damage the specimen. These grips clamp with no backlash.
Can it hold round specimens?
Yes. The round-specimen version uses V-jaws for Ø 4–12 mm.
How do the grips mount to the actuator and load cell?
An M12 x 1.75 threaded stud. Adapters are available for other threads.
Can I use it for metals or composites?
Yes. It also suits fatigue tests on steel, aluminum, CFRP and fiberglass composites within 30 kN and -40 to +280 °C.
Is a higher-capacity or high-temperature version available?
Special builds rated 50, 100 or 200 kN, and a 30 kN version for up to +600 °C, are available on request.