Amsler Double Shear Fixture – ASTM B769

Amsler-type double shear fixture for 10 mm aluminum alloy rod and bar to ASTM B769, with replaceable hardened dies and a 100 kN heavy version.

Brand: Universal Grip  |  Model: UGC-SF-020  |  ● Configured to order

  • Amsler-type double shear for aluminum alloy specimens per ASTM B769

  • Round 10 mm (0.39 in) specimen sheared across two planes by a central punch

  • Replaceable hardened cutting dies, Ø 25.4 x 13.5 mm with 10 mm bore

  • Hardened, nickel-plated steel dies and punch keep cutting edges sharp

  • 100 kN (about 22,480 lbf) heavy version for higher-strength alloys

  • Spare dies and punches available, so worn edges never compromise results

Standards: ASTM B769  ·  ASTM B565

Trusted by Global Companies

Amsler double shear fixture for aluminum alloys

The UGC-SF-020 is an Amsler-type fixture for double shear testing of aluminum alloy specimens to ASTM B769. Mills, extruders and aerospace suppliers use it to measure the shear strength of rod, bar and rivet wire, a property that tensile results alone do not give you. A standard version and a 100 kN (about 22,480 lbf) heavy version cover everyday alloys and the higher-strength tempers. See our complete ASTM B769 guide for specimen preparation, procedure and calculations.

 

How it works

A round specimen of 10 mm (0.39 in) diameter is passed through hardened cutting dies on either side and through a central punch. As load is applied, the punch shears the specimen across two planes at once. With two shear areas carrying the load, the fixture keeps the specimen in near-pure shear and gives the double shear strength the method calls for.

 

Dies, punch and versions

The cutting dies measure Ø 25.4 mm (1.00 in) by 13.5 mm (0.53 in) high, with a 10 mm bore. Dies and punch are made of hardened, nickel-plated steel. Their edges are what define the shear plane, so both are replaceable: when an edge starts to round, a new die set or punch restores the geometry instead of letting results drift. Order spares with the fixture if you run high volumes.

  • Standard fixture for 10 mm specimens, with a male clevis coupling with a Ø 31.8 mm (1-1/4 in) pin hole. Capacity is confirmed when we quote.

  • 100 kN fixture for higher-strength alloys, with a male Ø 50 mm coupling.

  • Replacement cutting dies (Ø 25.4 x 13.5 mm) and replacement punch.

 

Typical applications

  • Aluminum alloy rod and bar

  • Rivet wire

  • Aerospace alloys

  • Mill quality control and certification

Labs that also test rivet and cold-heading wire to ASTM B565 use the same double shear principle. Tell us your wire diameter and we will confirm whether the standard dies suit it or a special die bore is needed. For tension-mode double shear of wires and rivets to EN 3238, see the UGC-SF-030.

 

Machine fit

The 100 kN version suits a Galdabini 100 kN benchtop universal testing machine or a floor-standing frame. Lower-capacity frames can run the standard version within its confirmed rating. Couplings for other machines are available through our adapters. Tell us your specimen diameter, alloy and temper, and machine coupling, and we will configure the fixture and spare die sets: request a quote.

Product video

At a glance

  • Test type: Double shear of round specimens

  • Shear planes: Two

  • Max. load: 100 kN (22,480 lbf) heavy version; standard version on request

  • Specimen diameter: 10 mm (0.39 in)

  • Cutting dies: Ø 25.4 mm x 13.5 mm high, 10 mm bore

  • Punch: Central punch, replaceable

Amsler Double Shear Fixture Specifications

TEST TYPE

Double shear of round specimens

 

SHEAR PLANES

Two

 

MAX. LOAD

100 kN (22,480 lbf) heavy version; standard version on request

 

SPECIMEN DIAMETER

10 mm (0.39 in)

 

CUTTING DIES

Ø 25.4 mm x 13.5 mm high, 10 mm bore

 

PUNCH

Central punch, replaceable

 

COUPLING

Male clevis, Ø 31.8 mm (1-1/4 in) pin hole; 100 kN version: male Ø 50 mm

 

MATERIAL

Steel, hardened, nickel-plated

 

TEMPERATURE RANGE

Ambient

 

DRAWINGS

Dimensional drawings and 3D models on request

 

PART REF.

UGC-SF-020

 

Models & Options

  • Standard fixture — 10 mm specimens, male clevis with Ø 31.8 mm (1-1/4 in) pin hole (Specify at order)

  • 100 kN fixture — Heavy version for higher-strength alloys, male Ø 50 mm coupling (Specify at order)

  • Cutting dies — Replacement set, Ø 25.4 x 13.5 mm, 10 mm bore (Specify at order)

  • Punch — Replacement central punch (Specify at order)

  • Other specimen diameters — Special die bores for other rod or wire sizes (On request)

What's Included

  • Amsler double shear fixture with hardened cutting dies and central punch

  • Coupling per version: male clevis Ø 31.8 mm pin hole (standard) or male Ø 50 mm (100 kN)

  • Spare die sets and punches ordered separately as needed

Supported Test Standards

  • ASTM B769 — Shear testing of aluminum alloys (double shear)

  • ASTM B565 — Shear testing of aluminum rivets and cold-heading wire and rod – related; confirm specimen diameter

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-SF-030 Double Shear Fixture for Wires and Rivets EN 3238

UGC-SF-030 Double Shear Fixture for Wires and Rivets EN 3238

Tension-mode double shear fixture for wires and rivets to EN 3238, with a split housing, central blade and knurled clamp

UGC-SF-010 Single Shear Fastener Fixture NASM 1312-20

UGC-SF-010 Single Shear Fastener Fixture NASM 1312-20

50 kN single shear fixture for aerospace bolts, pins and rivets, built to the NASM 1312-20 and MIL-STD-1312-20 Fig. 3a a

UGC-SF-040 Arcan Shear Test Fixture

UGC-SF-040 Arcan Shear Test Fixture

Angle-adjustable Arcan fixture for pure shear and mixed-mode tests on composites, clear wood and bonded joints, in 20 kN

UGC-SF-050 Reinforcing Steel Weld Shear Fixture EN 10080

UGC-SF-050 Reinforcing Steel Weld Shear Fixture EN 10080

50 kN weld shear fixture for welded joints in reinforcing steel to EN 10080, with 6 mm and 9 mm inserts at test angles o

Amsler Double Shear Fixture FAQ

Is this the fixture for ASTM B769?

Yes. It is an Amsler-type double shear fixture built for ASTM B769 testing of aluminum alloys. Our ASTM B769 guide covers specimen preparation, procedure and calculations.

 

What specimen diameter does it take?

Round specimens of 10 mm (0.39 in) diameter, set by the 10 mm bore of the cutting dies. Other diameters need special dies; ask us.

 

What is the difference between the standard and the 100 kN version?

The 100 kN version is built for higher-strength alloys and uses a male Ø 50 mm coupling. The standard version uses a male clevis with a Ø 31.8 mm pin hole, and its capacity is confirmed at quote.

 

Why are the dies and punch replaceable?

The die and punch edges define the shear planes. Replacing a worn set restores sharp edges so results stay consistent.

 

Can I use it for rivet wire to ASTM B565?

The same double shear principle applies. Send us your wire diameter and we will confirm whether the standard dies suit it; see our ASTM B565 guide.

 

Which machine does it need?

For the 100 kN version, a Galdabini 100 kN benchtop machine or a floor-standing frame. Adapters are available for other couplings.

 

Need help configuring a shear fixture?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.