Arcan Shear Test Fixture – 20 kN and 50 kN Versions

Angle-adjustable Arcan fixture for pure shear and mixed-mode tests on composites, clear wood and bonded joints, in 20 kN and 50 kN versions.

Brand: Universal Grip  |  Model: UGC-SF-040  |  ● Configured to order

  • Loading angle adjustable from pure shear to combined shear-tension

  • 50 kN (about 11,240 lbf) versions for fiber-reinforced composites

  • 20 kN (about 4,500 lbf) version for shear properties of clear wood

  • Extended-temperature composite version for -60 to +120 °C (-76 to +248 °F)

  • Hardened, nickel-plated steel construction for long service life

  • Clevis couplings Ø 15.9 mm, Ø 31.8 mm or male Ø 35 mm; others via adapter

Downloads and standards

Trusted by Global Companies

Arcan fixture for shear and mixed-mode testing

The UGC-SF-040 is an Arcan-type fixture for determining the shear and combined shear-tension properties of fiber-reinforced composites, clear wood and adhesive joints. It is the tool of choice in research and material development labs that need more than a single shear number: by changing the loading angle, the same specimen geometry can be tested in pure shear, in mixed-mode, or anywhere in between, which makes it possible to map a failure envelope.

 

How it works

The specimen, typically an Arcan butterfly or block specimen, is clamped between two semicircular halves. Each half has a series of pin positions around its edge. Choosing which pair of holes connects to the machine sets the angle between the load line and the specimen's central section. One position gives pure shear across the specimen's waist; moving around the arc adds a tensile component, giving controlled mixed-mode loading. Specimen size is agreed at order to suit your material.

 

Versions

  • Composite Arcan, 50 kN: for fiber-reinforced composites, rated -60 to +120 °C (-76 to +248 °F), with a male Ø 35 mm coupling and 13 mm pin.

  • Angle-adjustable shear grip, 50 kN: rated 0 to +70 °C (32 to 158 °F), with a female clevis for a Ø 31.8 mm (1-1/4 in) pin.

  • Clear wood Arcan, 20 kN: for shear properties of clear wood, with a female clevis for a Ø 15.9 mm (5/8 in) pin.

All versions are made of hardened, nickel-plated steel, so the pin holes and clamping faces keep their geometry across repeated angle changes. Other couplings are available through adapters.

 

Typical applications

  • Fiber-reinforced composites

  • Clear wood

  • Adhesive joints

  • Mixed-mode and failure-envelope research

 

Choosing between Arcan and V-notch methods

If your specification names a standardized composite shear method, a dedicated fixture is usually the simpler path. The UGC-SF-230 V-notched beam fixture is built for ASTM D5379, and the UGC-SF-250 V-notched rail fixture for ASTM D7078. The Arcan fixture earns its place when you need several loading angles from one setup.

 

Machine fit

The 20 kN wood version runs on a Galdabini 25 kN benchtop machine, the 50 kN versions on a 50 kN benchtop machine. For the -60 to +120 °C composite version, pair it with an environmental chamber. Tell us your material, specimen geometry, required capacity, temperature range and machine coupling, and we will configure the fixture: request a quote.

Product video

At a glance

  • Test type: Arcan shear / mixed-mode

  • Max. load: 20 kN (wood) or 50 kN (composites)

  • Loading angle: Adjustable via pin positions

  • Specimen: Arcan butterfly or block specimen (size on request)

  • Coupling: Female clevis Ø 15.9 mm or Ø 31.8 mm; male Ø 35 mm (13 mm pin); others via adapter

  • Material: Steel, hardened, nickel-plated

Arcan Shear Test Fixture Specifications

TEST TYPE

Arcan shear / mixed-mode

 

MAX. LOAD

20 kN (wood) or 50 kN (composites)

 

LOADING ANGLE

Adjustable via pin positions

 

SPECIMEN

Arcan butterfly or block specimen (size on request)

 

COUPLING

Female clevis Ø 15.9 mm or Ø 31.8 mm; male Ø 35 mm (13 mm pin); others via adapter

 

MATERIAL

Steel, hardened, nickel-plated

 

TEMPERATURE RANGE

0 … +70 °C; -60 … +120 °C version

 

DRAWINGS

Dimensional drawings and 3D models on request

 

PART REF.

UGC-SF-040

 

Models & Options

  • Composite Arcan, 50 kN — -60 … +120 °C, male clevis Ø 35 mm with 13 mm pin (Specify at order)

  • Angle-adjustable shear grip, 50 kN — 0 … +70 °C, female clevis Ø 31.8 mm (1-1/4 in) (Specify at order)

  • Clear wood Arcan, 20 kN — Female clevis Ø 15.9 mm (5/8 in) (Specify at order)

  • Coupling adapters — For other machine connections (On request)

What's Included

  • Arcan fixture with two semicircular halves and pin positions for angle adjustment

  • Coupling per chosen version

  • Specimen size agreed at order; specimens are not included

Supported Test Standards

Standards list

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-SF-230 V-Notched Beam Shear Fixture

UGC-SF-230 V-Notched Beam Shear Fixture

Iosipescu V-notched beam shear fixture for composite laminates, rated to 50 kN, with the compression probe and specimen

UGC-SF-250 V-Notched Rail Shear Fixture

UGC-SF-250 V-Notched Rail Shear Fixture

Pair of V-notched rail shear grips for composites to ASTM D7078, in 50 kN and 100 kN versions plus a -100 to +600 °C hig

UGC-SF-030 Double Shear Fixture for Wires and Rivets EN 3238

UGC-SF-030 Double Shear Fixture for Wires and Rivets EN 3238

Tension-mode double shear fixture for wires and rivets to EN 3238, with a split housing, central blade and knurled clamp

UGC-SF-050 Reinforcing Steel Weld Shear Fixture EN 10080

UGC-SF-050 Reinforcing Steel Weld Shear Fixture EN 10080

50 kN weld shear fixture for welded joints in reinforcing steel to EN 10080, with 6 mm and 9 mm inserts at test angles o

Arcan Shear Test Fixture FAQ

What is an Arcan fixture used for?

Measuring shear and combined shear-tension properties. Changing the pin position changes the loading angle, from pure shear to mixed-mode.

 

Which version should I choose?

The 20 kN version is for clear wood. The two 50 kN versions are for composites; pick the -60 … +120 °C version if you test outside 0 … +70 °C.

 

What specimen does it take?

An Arcan butterfly or block specimen. Specimen size is agreed at order to suit your material and thickness.

 

Can it test at low or high temperature?

The composite version is rated -60 to +120 °C (-76 to +248 °F) and is used with an environmental chamber. The other versions are rated 0 to +70 °C.

 

Does it replace an ASTM D5379 or D7078 fixture?

Not when your specification names one of those methods. Use the UGC-SF-230 for D5379 or the UGC-SF-250 for D7078.

 

Which couplings are available?

Female clevis Ø 15.9 mm or Ø 31.8 mm, or male Ø 35 mm with a 13 mm pin, depending on version. Adapters cover other machines.

 

Need help configuring a shear fixture?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.

bottom of page