Home › Products › Video Extensometers › UGC-VE-4116-D 16 MPx Dual-Camera Video Extensometer





UGC-VE-4116-D 16 MPx Dual-Camera Video Extensometer
Two 16 MPx cameras in a single 400 mm housing for a larger joined field of view. The longest field of view in its series.
Brand: Universal Grip | Model: UGC-VE-4116-D | ● Configured to order
ISO 9513 Class 0.5 measuring length 625* mm; Class 1 950* mm
70 Hz typical sampling, up to 1 kHz with a narrowed view
Two cameras in one housing — joined field of view, one unit to mount
All-in-one: camera, C-mount lens, blue LED light and USB light relay
Non-contact: stays on through break, no knife edges
Live output to your test machine (digital, analog, pulse) and API control
Downloads: Video Extensometer Inquiry Form (Excel)
Standards: ASTM D4595 · ASTM D2256 · ASTM D885 · ASTM A615 · ASTM D4632
Trusted by Global Companies


























Two 16 MPx cameras in a single 400 mm housing: one box to mount and cable
The UGC-VE-4116-D is a non-contact optical (video) extensometer from the X-Sight 4000-D Series, supplied, configured and supported by Universal Grip. It tracks gauge marks on the specimen with two 16 MPx cameras in one housing and reports strain, total length, delta length and angle in real time, so it can stay on the specimen through break with no knife edges to slip or damage. Two cameras share one housing and one 400 mm LED, joined with a calibration grid for long or highly elongating specimens — with a single box to mount and cable.
Measuring length, accuracy class and sampling rate
Per ISO 9513 the UGC-VE-4116-D measures 625* mm at Class 0.5, 950* mm at Class 1 and 1610* mm (* additional light may be needed) at Class 2. Class 0.5 is the tight class labs choose for modulus and proof-strength work; ISO 6892-1 calls for Class 1 or better for yield and proof strength and allows Class 2 for elongation after fracture.
Typical sampling is 70 Hz (23 Hz with the full image), and narrowing the view width raises it to 1 kHz. Higher pixel counts trade frame rate for field of view: at this resolution the camera favours reach and resolution over raw speed.
As a sizing check: the Class 2 measuring length of 1610 mm (extra lighting may be needed at this length) is long enough to follow a gauge section of about 1.6 m, to measure several gauge lengths on one long specimen, or to watch more than one specimen at once. We confirm the usable field from your specimen, grips and expected elongation.
What the UGC-VE-4116-D is best for
wide-width geotextiles and geogrids to ASTM D4595
full-size rebar and structural specimens to ASTM A615
rope, cable and webbing
large components, assemblies and several specimens in one view
Two cameras, one housing
The UGC-VE-4116-D puts two cameras and one 400 mm LED in a single 424 mm housing. The cameras sit up to 310 mm apart and are joined with a calibration grid for a longer field of view, but there is only one box to mount, align and cable. That keeps setup fast on frames where space beside the load string is tight.
Choosing the lens and working distance
Measuring length, lens focal length and working distance are linked: choose two and the third follows. For the Class 0.5 field of view (625* mm high by 188 mm wide) the working distances are 12 mm lens 262 mm, 16 mm lens 352 mm, 25 mm lens 556 mm, 35 mm lens 782 mm, 50 mm lens 1,122 mm (from the front cover). Longer working distances reduce LED intensity and can need longer shutter times, so tell us your frame and chamber geometry and we will pick the lens.
Connects to your testing machine
Strain is streamed live to the test-machine controller over DOLI Binary, MODBUS, HP Video, TCP/IP or RS232, as an analog signal through auxiliary D/A converters, or as an encoder-style quadrature pulse. The JSON and MRT APIs let your test software start and stop measurement, switch methods and set the gauge length. That is how we retrofit video extensometers to Instron, MTS, ZwickRoell, Shimadzu, Test Resources, ADMET and Galdabini frames; see retrofitting a video extensometer to an existing machine.
Compare the UGC-VE-4116-D with similar models
UGC-VE-4109-D: 4000-D Series dual-camera, 9 MPx, Class 0.5 420 mm, 75 Hz typical
UGC-VE-4106-D: 4000-D Series dual-camera, ASTM-optimized, ASTM E83 Class B-1, 200 Hz typical
UGC-VE-2116: 2000 Series single-camera, 16 MPx, Class 0.5 330 mm, 70 Hz typical
UGC-VE-2216: 2000 Series dual-camera, 16 MPx, Class 0.5 640 mm, 70 Hz typical
Not sure which model fits? Read how to choose an extensometer, gauge length selection for video extensometers, or compare all models on our video extensometer page.
Video Extensometer Sample Marking Guide
At a glance
Cameras: 2 × 16 MPx industrial cameras in one housing
Sampling rate: 70 Hz typical · 23 Hz at full view · up to 1 kHz with reduced view width
Measuring length (ISO 9513): Class 0.5: 625* mm · Class 1: 950* mm · Class 2: 1610* mm (* additional light may be needed)
Camera span: Up to 310 mm between cameras; add a second unit for larger fields of view
Field of view, Class 0.5 (H × W): 625* × 188 mm
Field of view, Class 1 (H × W): 950* × 376 mm
UGC-VE-4116-D 16 MPx Dual-Camera Video Extensometer Specifications
MODEL
UGC-VE-4116-D · 4000-D Series, dual-camera
CAMERAS
2 × 16 MPx industrial cameras in one housing
SAMPLING RATE
70 Hz typical · 23 Hz at full view · up to 1 kHz with reduced view width
MEASURING LENGTH (ISO 9513)
Class 0.5: 625* mm · Class 1: 950* mm · Class 2: 1610* mm (* additional light may be needed)
CAMERA SPAN
Up to 310 mm between cameras; add a second unit for larger fields of view
FIELD OF VIEW, CLASS 0.5 (H × W)
625* × 188 mm
FIELD OF VIEW, CLASS 1 (H × W)
950* × 376 mm
FIELD OF VIEW, CLASS 2 (H × W)
1610* × 752 mm
WORKING DISTANCE
300–500 mm typical from front cover; extendable on request
LENS
2 × C-mount; 12, 16, 25, 35 or 50 mm focal length selected per application
LIGHT
400 mm blue LED (465 nm, 330 lm, 16 W), USB-switched
PC INTERFACE
USB 3.0 per camera (2 total) + USB 2.0 for light relay; 3 m cables standard
DATA OUTPUT
DOLI Binary, MODBUS, HP Video, TCP/IP, RS232; analog via D/A; quadrature pulse; JSON and MRT API
HOUSING DIMENSIONS
424 × 187 × 80 mm (L × W × H)
WEIGHT
2.6 kg
POWER
24 V DC via RJ45 power cable (802.3af Mode B); 23 W
MOUNTING
Tripod 1/4"-20 UNC (center) or 2 × M6 on a 165 mm vertical span
ENVIRONMENT
Indoor; 5–40 °C, 30–70 % RH
SOFTWARE
Windows 10/11 64-bit or Windows Server 2019/2022; perpetual license on USB key
PC REQUIREMENTS
2 GHz dual-core min. (4 GHz, 8+ cores rec.); 4 GB RAM min. (16 GB rec.); OpenGL 3.0
STANDARDS
ISO 9513 accuracy classes; used for ISO 6892-1, ASTM E8 and similar methods
Models & Options
C-mount lens — 12, 16, 25, 35 or 50 mm focal length chosen from your gauge length and working distance (Specify at order)
AX module (included) — Point, Line and Extreme Line probes for axial strain and elongation (AX)
Transverse module — Transverse line and bend line for r-value, Poisson's ratio and deflection (TR)
Torsion / crack module — Torsion line and crack-opening probe (TO / CR)
Rebar module — Rebar line for reinforcing-steel testing (ITT)
DIC / post-processing — DIC area strain mapping and re-analysis of recorded tests (DIC / PP)
High-speed, device input, color — High-speed recorder, external force/temperature/pressure input, color value distribution (HS / DI / LVD)
Accessories — Calibration grids, tripods, polarization kits, PCIe USB 3.0 card, pulse generator, D/A converters (On request)
Sample marking — Gauge-mark marking tools for fast, repeatable marks (Marking tool)
What's Included
VE unit with 2 cameras, 2 lenses, 400 mm LED & built-in USB relay
Joined calibration grid sized to the application
Cable harness, 24 V power supply
Software on installation USB + USB license key
Axial software module (perpetual license)
1 year software updates & technical support
Supported Test Standards
ASTM D4595 — Wide-width tensile test of geotextiles
ASTM D2256 — Tensile properties of yarns
ASTM D885 — Tire cords and industrial filament yarns
ASTM A615 — Deformed carbon-steel bars for concrete reinforcement
ASTM D4632 — Grab breaking load of geotextiles
Compatible Grips & Fixtures
Universal testing machines — Retrofit to Instron, MTS, ZwickRoell, Shimadzu, Test Resources, ADMET or Galdabini frames
Tensile grips — Wedge, vise, roller and pneumatic grips; the camera views the specimen between them
Environmental chambers — Measure through the chamber window with the right lighting
Marking tool — Repeatable gauge marks for video extensometers
Clip-on extensometers — Contact alternative for single-gauge-length tests
All extensometers — Clip-on, long-travel, video and 3D extensometers
“Customer quote”
Related Equipment
UGC-VE-4109-D 9 MPx Dual-Camera
4000-D Series dual-camera, 9 MPx
UGC-VE-4106-D ASTM Dual-Camera
4000-D Series dual-camera, ASTM-optimized
UGC-VE-2116 16 MPx Single-Camera
2000 Series single-camera, 16 MPx
UGC-VE-2216 16 MPx Dual-Camera
2000 Series dual-camera, 16 MPx
UGC-VE-4116-D 16 MPx Dual-Camera Video Extensometer FAQ
What gauge length and field of view does the UGC-VE-4116-D cover?
Per ISO 9513, Class 0.5: 625* mm; Class 1: 950* mm; Class 2: 1610* mm (* additional light may be needed). The exact field depends on the lens and working distance, which we pick from your specimen and frame.
Which ISO 9513 class do I need?
ISO 6892-1 calls for Class 1 or better for yield and proof strength and allows Class 2 for elongation after fracture. Many labs use Class 0.5 for modulus. Plastics methods such as ISO 527 set their own accuracy requirements.
How fast does the UGC-VE-4116-D sample?
70 Hz typical, 23 Hz with the full image, and up to 1 kHz with a narrowed view width. Data streams to the test machine controller in real time.
Will the UGC-VE-4116-D work with my existing testing machine?
Yes, in most cases. It outputs DOLI Binary, MODBUS, HP Video, TCP/IP, RS232, analog via D/A converters or a quadrature pulse, and the JSON/MRT APIs allow remote start/stop and gauge-length setting. Tell us the make and model of your frame and controller. See retrofitting a video extensometer.
What does the -D in UGC-VE-4116-D mean?
Two cameras in one 400 mm housing. The cameras sit up to 310 mm apart and are joined with a calibration grid, giving a longer field of view with only one unit to mount and cable.
What software comes with the UGC-VE-4116-D?
The axial (AX) module with Point, Line and Extreme Line probes, on a perpetual license stored on a USB key, with one year of updates and technical support. Optional modules add transverse strain (r-value, Poisson's ratio), torsion, crack opening, rebar, DIC mapping and post-processing.
How do I mark specimens for a video extensometer?
Use high-contrast dots or lines at the gauge length. Our marking tools and the sample marking guide cover the common methods.
How do I get a quote for the UGC-VE-4116-D?
Send your material, test standard, specimen size and gauge length, expected elongation, and your testing machine make/model. Use the Request a Quote button, call 978-882-1480, or download the video extensometer inquiry form.



