top of page

Home  ›  Products  ›  Video Extensometers  ›  UGC-VE-4116 16 MPx Single-Camera Video Extensometer

UGC-VE-4116 16 MPx 4000 Series single-camera video extensometer by Universal Grip
UGC-VE 4000 Series video extensometer with 400 mm LED
Universal Grip video extensometer tracking a tensile specimen
Video extensometer strain map on tensile specimens
Optical extensometer software kit

UGC-VE-4116 16 MPx Single-Camera Video Extensometer

16 MPx non-contact optical extensometer in a 400 mm housing with a 400 mm LED light. Maximum field of view for long specimens.

Brand: Universal Grip  |  Model: UGC-VE-4116  |  ● Configured to order

  • ISO 9513 Class 0.5 measuring length 330 mm; Class 1 660 mm

  • 70 Hz typical sampling, up to 1 kHz with a narrowed view

  • All-in-one: camera, C-mount lens, blue LED light and USB light relay

  • 400 mm LED for long or high-elongation specimens; stackable for joined fields of view

  • Non-contact: stays on through break, no knife edges

  • Live output to your test machine (digital, analog, pulse) and API control

Trusted by Global Companies

16 MPx video extensometer with a 400 mm LED for a larger illuminated area

The UGC-VE-4116 is a non-contact optical (video) extensometer from the X-Sight 4000 Series, supplied, configured and supported by Universal Grip. It tracks gauge marks on the specimen with one 16 MPx camera and reports strain, total length, delta length and angle in real time, so it can stay on the specimen through break with no knife edges to slip or damage. The longer housing and 400 mm LED illuminate a larger area than the 2000 Series — the better choice for long or high-elongation specimens.

 

Measuring length, accuracy class and sampling rate

Per ISO 9513 the UGC-VE-4116 measures 330 mm at Class 0.5, 660 mm at Class 1 and 1320 mm at Class 2. Class 0.5 is the tight class labs choose for modulus and proof-strength work; ISO 6892-1 calls for Class 1 or better for yield and proof strength and allows Class 2 for elongation after fracture.

Typical sampling is 70 Hz (23 Hz with the full image), and narrowing the view width raises it to 1 kHz. Higher pixel counts trade frame rate for field of view: at this resolution the camera favours reach and resolution over raw speed.

As a sizing check: the Class 2 measuring length of 1320 mm is long enough to follow a gauge section of about 1.3 m, to measure several gauge lengths on one long specimen, or to watch more than one specimen at once. We confirm the usable field from your specimen, grips and expected elongation.

 

What the UGC-VE-4116 is best for

  • wide-width geotextile tests to ASTM D4595 and geogrids

  • rope, cord and yarn to ASTM D2256 and ASTM D885

  • rebar and long structural specimens to ASTM A615

  • large components and several specimens in one field of view

 

When to choose the 400 mm housing

The 4000 Series uses the same 16 MPx camera as the UGC-VE-2116 but a 424 mm housing with a 400 mm, 330 lm LED. The longer light evenly illuminates a larger area, which helps with long or high-elongation specimens, darker or reflective surfaces and viewing through a temperature-chamber window. If your field of view is short, the more compact UGC-VE-2116 is usually enough.

 

Choosing the lens and working distance

Measuring length, lens focal length and working distance are linked: choose two and the third follows. For the Class 0.5 field of view (330 mm high by 188 mm wide) the working distances are 12 mm lens 262 mm, 16 mm lens 352 mm, 25 mm lens 556 mm, 35 mm lens 782 mm, 50 mm lens 1,122 mm (from the front cover). Longer working distances reduce LED intensity and can need longer shutter times, so tell us your frame and chamber geometry and we will pick the lens.

 

Connects to your testing machine

Strain is streamed live to the test-machine controller over DOLI Binary, MODBUS, HP Video, TCP/IP or RS232, as an analog signal through auxiliary D/A converters, or as an encoder-style quadrature pulse. The JSON and MRT APIs let your test software start and stop measurement, switch methods and set the gauge length. That is how we retrofit video extensometers to Instron, MTS, ZwickRoell, Shimadzu, Test Resources, ADMET and Galdabini frames; see retrofitting a video extensometer to an existing machine.

 

Compare the UGC-VE-4116 with similar models

  • UGC-VE-4109: 4000 Series single-camera, 9 MPx, Class 0.5 220 mm, 75 Hz typical

  • UGC-VE-4106: 4000 Series single-camera, ASTM-optimized, ASTM E83 Class B-1, 200 Hz typical

  • UGC-VE-2116: 2000 Series single-camera, 16 MPx, Class 0.5 330 mm, 70 Hz typical

  • UGC-VE-2216: 2000 Series dual-camera, 16 MPx, Class 0.5 640 mm, 70 Hz typical

Not sure which model fits? Read how to choose an extensometer, gauge length selection for video extensometers, or compare all models on our video extensometer page.

Video Extensometer vs. Clip-On Extensometers

At a glance

  • Cameras: 1 × 16 MPx industrial camera

  • Sampling rate: 70 Hz typical · 23 Hz at full view · up to 1 kHz with reduced view width

  • Measuring length (ISO 9513): Class 0.5: 330 mm · Class 1: 660 mm · Class 2: 1320 mm

  • Field of view, Class 0.5 (H × W): 330 × 188 mm

  • Field of view, Class 1 (H × W): 660 × 376 mm

  • Field of view, Class 2 (H × W): 1320 × 752 mm

UGC-VE-4116 16 MPx Single-Camera Video Extensometer Specifications

MODEL

UGC-VE-4116 · 4000 Series, single-camera

 

CAMERAS

1 × 16 MPx industrial camera

 

SAMPLING RATE

70 Hz typical · 23 Hz at full view · up to 1 kHz with reduced view width

 

MEASURING LENGTH (ISO 9513)

Class 0.5: 330 mm · Class 1: 660 mm · Class 2: 1320 mm

 

FIELD OF VIEW, CLASS 0.5 (H × W)

330 × 188 mm

 

FIELD OF VIEW, CLASS 1 (H × W)

660 × 376 mm

 

FIELD OF VIEW, CLASS 2 (H × W)

1320 × 752 mm

 

WORKING DISTANCE

300–500 mm typical from front cover; extendable on request

 

LENS

C-mount; 12, 16, 25, 35 or 50 mm focal length selected per application

 

LIGHT

400 mm blue LED (465 nm, 330 lm, 16 W), USB-switched

 

PC INTERFACE

USB 3.0 per camera (1 total) + USB 2.0 for light relay; 3 m cables standard

 

DATA OUTPUT

DOLI Binary, MODBUS, HP Video, TCP/IP, RS232; analog via D/A; quadrature pulse; JSON and MRT API

 

HOUSING DIMENSIONS

424 × 187 × 80 mm (L × W × H)

 

WEIGHT

2.5 kg

 

POWER

24 V DC via RJ45 power cable (802.3af Mode B); 20 W

 

MOUNTING

Tripod 1/4"-20 UNC (center) or 2 × M6 on a 165 mm vertical span

 

ENVIRONMENT

Indoor; 5–40 °C, 30–70 % RH

 

SOFTWARE

Windows 10/11 64-bit or Windows Server 2019/2022; perpetual license on USB key

 

PC REQUIREMENTS

2 GHz dual-core min. (4 GHz, 8+ cores rec.); 4 GB RAM min. (16 GB rec.); OpenGL 3.0

 

STANDARDS

ISO 9513 accuracy classes; used for ISO 6892-1, ASTM E8 and similar methods

 

Models & Options

  • C-mount lens — 12, 16, 25, 35 or 50 mm focal length chosen from your gauge length and working distance (Specify at order)

  • AX module (included) — Point, Line and Extreme Line probes for axial strain and elongation (AX)

  • Transverse module — Transverse line and bend line for r-value, Poisson's ratio and deflection (TR)

  • Torsion / crack module — Torsion line and crack-opening probe (TO / CR)

  • Rebar module — Rebar line for reinforcing-steel testing (ITT)

  • DIC / post-processing — DIC area strain mapping and re-analysis of recorded tests (DIC / PP)

  • High-speed, device input, color — High-speed recorder, external force/temperature/pressure input, color value distribution (HS / DI / LVD)

  • Accessories — Calibration grids, tripods, polarization kits, PCIe USB 3.0 card, pulse generator, D/A converters (On request)

  • Sample marking — Gauge-mark marking tools for fast, repeatable marks (Marking tool)

  • Upgrade path — Add a second unit for the 4200 two-unit configuration (4200)

What's Included

  • VE unit with lens, 400 mm LED & built-in USB relay

  • Mono calibration grid sized to the application

  • Cable harness, 24 V power supply

  • Software on installation USB + USB license key

  • Axial software module (perpetual license)

  • 1 year software updates & technical support

Supported Test Standards

  • ASTM D4595 — Wide-width tensile test of geotextiles

  • ASTM D2256 — Tensile properties of yarns

  • ASTM D885 — Tire cords and industrial filament yarns

  • ASTM A615 — Deformed carbon-steel bars for concrete reinforcement

  • ASTM D4632 — Grab breaking load of geotextiles

Compatible Grips & Fixtures

“Customer quote”

Related Equipment

UGC-VE-4109 9 MPx Single-Camera

UGC-VE-4109 9 MPx Single-Camera

4000 Series single-camera, 9 MPx

UGC-VE-4106 ASTM

UGC-VE-4106 ASTM

4000 Series single-camera, ASTM-optimized

UGC-VE-2116 16 MPx Single-Camera

UGC-VE-2116 16 MPx Single-Camera

2000 Series single-camera, 16 MPx

UGC-VE-2216 16 MPx Dual-Camera

UGC-VE-2216 16 MPx Dual-Camera

2000 Series dual-camera, 16 MPx

UGC-VE-4116 16 MPx Single-Camera Video Extensometer FAQ

What gauge length and field of view does the UGC-VE-4116 cover?

Per ISO 9513, Class 0.5: 330 mm; Class 1: 660 mm; Class 2: 1320 mm. The exact field depends on the lens and working distance, which we pick from your specimen and frame.

 

Which ISO 9513 class do I need?

ISO 6892-1 calls for Class 1 or better for yield and proof strength and allows Class 2 for elongation after fracture. Many labs use Class 0.5 for modulus. Plastics methods such as ISO 527 set their own accuracy requirements.

 

How fast does the UGC-VE-4116 sample?

70 Hz typical, 23 Hz with the full image, and up to 1 kHz with a narrowed view width. Data streams to the test machine controller in real time.

 

Will the UGC-VE-4116 work with my existing testing machine?

Yes, in most cases. It outputs DOLI Binary, MODBUS, HP Video, TCP/IP, RS232, analog via D/A converters or a quadrature pulse, and the JSON/MRT APIs allow remote start/stop and gauge-length setting. Tell us the make and model of your frame and controller. See retrofitting a video extensometer.

 

Should I choose the UGC-VE-4116 or the UGC-VE-2116?

Both use the same camera. The UGC-VE-4116 adds a 400 mm, 330 lm LED that evenly lights a larger area for long or high-elongation specimens and chamber windows. If your field of view is short, the compact <a href="/equipment/ugc-ve-2116-video-extensometer">UGC-VE-2116</a> is usually enough.

 

What software comes with the UGC-VE-4116?

The axial (AX) module with Point, Line and Extreme Line probes, on a perpetual license stored on a USB key, with one year of updates and technical support. Optional modules add transverse strain (r-value, Poisson's ratio), torsion, crack opening, rebar, DIC mapping and post-processing.

 

How do I mark specimens for a video extensometer?

Use high-contrast dots or lines at the gauge length. Our marking tools and the sample marking guide cover the common methods.

 

How do I get a quote for the UGC-VE-4116?

Send your material, test standard, specimen size and gauge length, expected elongation, and your testing machine make/model. Use the Request a Quote button, call 978-882-1480, or download the video extensometer inquiry form.

 

Need help configuring a video extensometer?

Tell us your material, test standard and specimen size. We will recommend the frame, force sensor, grips and fixtures, and send a quote.

bottom of page